新生儿高频呼吸机环境噪音产生量的台架研究
该台架研究比较了Sensormedics 3100A、Bunnell Life Pulse 203和204及Dräger VN500在多种高频通气设置下的噪音水平。Dräger VN500整体最安静,平均声级49.8±0.49 dB,其余三款均超过53.6 dB;增加振幅/峰值吸气压力会显著增加所有呼吸机的噪音。
要点
- Dräger VN500在高频模式下产生的噪音显著低于Sensormedics和Bunnell呼吸机。
- 增加振幅/峰值吸气压力会显著增加所有受试呼吸机的噪音。
- 调节频率/速率和平均气道压/呼气末正压对噪音影响极小。
核心结论
- Dräger VN500在高频模式下产生的噪音显著低于Sensormedics和Bunnell呼吸机,可能减少呼吸机噪音相关不良结局的潜在风险。
研究设计与样本
- 台架研究,比较Sensormedics 3100A、Bunnell Life Pulse 203和204及Dräger VN500在高频模式下的噪音产生,涵盖广泛通气设置,所有测量重复三次,使用高保真声级计,采用方差分析和回归分析。
主要发现
- Dräger呼吸机整体最安静,所有设置下平均声级为49.8±0.49 dB。
- Sensormedics平均噪音为53.6±2.01 dB,Bunnell 203为54.1±1.09 dB,Bunnell 204为53.7±1.45 dB。
- 调节频率/速率和平均气道压/呼气末正压对噪音影响极小,增加振幅/峰值吸气压力导致所有呼吸机噪音显著增加。
- 在所有设置下Sensormedics和Bunnell均比Dräger更吵,且随振幅/峰值吸气压力增加差异更大。
临床怎么用
- 在选择高频呼吸机时,可考虑Dräger VN500以降低噪音暴露,尤其需较高振幅/峰值吸气压力时。
- 临床调节振幅/峰值吸气压力时需意识到噪音可能随之增加。
局限性
- 本研究为台架研究,未在临床环境中验证,也未评估噪音对患者结局的直接影响。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To assess sound levels of 4 high-frequency neonatal ventilators to determine whether there is a safety benefit in using modern high-frequency ventilators compared with older models. STUDY DESIGN: We performed a bench study comparing noise production of the Sensormedics 3100A Oscillator, Bunnell Life Pulse Jet Ventilators Model 203 and Model 204, and Dräger VN500 in high-frequency mode. A wide range of ventilation settings was examined. All measurements were performed in triplicate using a high-fidelity sound meter, with data analyzed using ANOVA and regression analyses. RESULTS: The Dräger ventilator was quietest overall, with average sound levels of 49.8 ± 0.49 dB across all settings. The average noise from the Sensormedics was 53.6 ± 2.01 dB, for Bunnell Model 203 was 54.1 ± 1.09 dB, and for Bunnell Model 204 was 53.7 ± 1.45 dB. Adjustments made to frequency/rate and mean airway pressure/positive end-expiratory pressure had minimal effect on noise, and increasing amplitude/peak inspiratory pressure resulted in significantly more noise by all ventilators. At all settings, the Sensormedics and Bunnell ventilators were louder than the Dräger, and the difference became greater as amplitude/peak inspiratory pressure was increased. CONCLUSIONS: The Dräger VN500 in high-frequency mode produces significantly less noise that both the Sensormedics and Bunnell ventilators. These data suggest that using the Dräger VN500 as a high-frequency ventilator may reduce the potential for adverse outcomes created by ventilator noise.