住院数据驱动生命体征参数减少儿科病房报警负担
单中心质量改进项目在20张床的儿科心血管急性病房实施基于年龄的住院数据驱动心率和呼吸频率参数,历史对照显示每监护床日总报警中位数从52降至41(下降21%,P < .001),期间无心肺骤停事件。
要点
- 实施数据驱动心率与呼吸频率参数后,每监护床日总报警中位数下降21%。
- 呼吸频率报警下降53%,心率报警下降17%。
- 实施后发生57次快速反应团队启动、8次急性呼吸代码事件、无心肺骤停事件。
核心结论
- 实施数据驱动的心率和迭代推导的呼吸频率参数可安全地将儿科急性病房总报警频率降低21%。
研究设计与样本
- 单中心质量改进项目,采用历史对照,在主要收治儿科心血管患者的20张床急性病房实施基于年龄的住院数据驱动心率和呼吸频率参数。
主要发现
- 每监护床日所有心肺监护报警中位数从基线52降至干预后41(P < .001)。
- 心率报警中位数从每监护床日9降至7.5(下降17%),呼吸频率报警从16.8降至8.0(下降53%)。
- 实施新参数后发生57次快速反应团队启动、8次急性呼吸代码事件、无心肺骤停事件。
- 事件发生时记录的心率和呼吸频率评估显示,所有超出旧默认参数范围的患者也超出新参数范围。
关键数值与推荐
- 主要结局为每监护床日报警数,目标为降低该数值。
- 平衡指标包括漏报快速反应团队启动、急性呼吸代码事件和心肺骤停事件的频率。
临床怎么用
- 在儿科急性病房可采用基于年龄的住院数据驱动心率和呼吸频率参数替代默认参数,以降低报警负担且不增加漏报风险。
局限性
- 单中心质量改进项目,采用历史对照,可能受时间趋势和未测量混杂因素影响。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: To implement data-driven vital sign parameters to reduce bedside monitor alarm burden. METHODS: Single-center, quality-improvement initiative with historical controls assessing the impact of age-based, inpatient-derived heart rate (HR) and respiratory rate (RR) parameters on a 20-bed acute care ward that serves primarily pediatric cardiology patients. The primary outcome was the number of alarms per monitored bed day (MBD) with the aim to decrease the alarms per MBD. Balancing measures included the frequency of missed rapid response team activations, acute respiratory code events, and cardiorespiratory arrest events in the unit with the new vital sign parameters. RESULTS: The median number of all cardiorespiratory monitor alarms per MBD decreased by 21% from 52 (baseline period) to 41 (postintervention period) (P < .001). This included a 17% decrease in the median HR alarms (9-7.5 per MBD) and a 53% drop in RR alarms (16.8-8.0 per MBD). There were 57 rapid response team activations, 8 acute respiratory code events, and no cardiorespiratory arrest events after the implementation of the new parameters. An evaluation of HRs and RRs recorded at the time of the event revealed that all patients with HRs and/or RRs out of range per former default parameters would also be out of range with the new parameters. CONCLUSIONS: Implementation of data-driven HR and iteratively derived RR parameters safely decreased the total alarm frequency by 21% in a pediatric acute care unit.
引用
- Kipps AK, Poole SF, Slaney C, et al. Inpatient-Derived Vital Sign Parameters Implementation: An Initiative to Decrease Alarm Burden. Pediatrics. 2017. PMID: 28687637. doi:10.1542/peds.2016-2458
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。