单心室先天性心脏病儿童脑组织氧合的无创光学定量
本研究纳入24例拟行心导管检查的单心室先天性心脏病儿童,比较频域近红外光谱(FD-NIRS)与连续波近红外光谱(CW-NIRS)及脉搏血氧仪对脑组织氧饱和度(StO2)和动脉血氧饱和度(SaO2)的测量准确性。结果显示FD-NIRS的StO2与侵入性静脉血氧饱和度(SvO2)强相关(R=0.74,P<.001),而CW-NIRS相关性弱且不显著(R=0.29,P=.34);FD-NIRS估测的SaO2与侵入性SaO2相关(R=0.84,P<.001),脉搏血氧仪在低饱和度时误差较大。FD-NIRS计算的氧摄取分数(OEF)与侵入性OEF强相关(R=0.61,P=.01)。
要点
- FD-NIRS测得的StO2与侵入性SvO2强相关(R=0.74),优于CW-NIRS(R=0.29,不显著)。
- FD-NIRS估测SaO2与侵入性SaO2相关(R=0.84),脉搏血氧仪在低饱和度时误差更大。
- FD-NIRS可独立量化OEF,与侵入性OEF强相关(R=0.61)。
核心结论
- FD-NIRS对单心室先天性心脏病患者StO2和SaO2的定量比CW-NIRS和脉搏血氧仪更准确。
- FD-NIRS可独立量化OEF,并与侵入性OEF强相关。
研究设计与样本
- 研究纳入24例拟行心导管检查的单心室先天性心脏病儿童。
- 术中在额头双侧进行FD-NIRS和CW-NIRS的StO2测量,同时用FD-NIRS和脉搏血氧仪获取SaO2。
- 将无创StO2和SaO2与侵入性上腔静脉(SVC)及降主动脉饱和度测量进行比较,并计算脑组织氧摄取分数(OEF)。
主要发现
- FD-NIRS StO2与侵入性SvO2强相关(R=0.74,P<.001;偏差-2.8%)。
- CW-NIRS StO2与侵入性SvO2相关性弱且不显著(R=0.29,P=.34;偏差-2.5%)。
- 侵入性SaO2与FD-NIRS(R=0.84,P<.001;偏差-3.2%)和脉搏血氧仪(R=0.68,P=.001;偏差0.2%)的SaO2指标相关,但脉搏血氧仪在低饱和度时误差更大。
- FD-NIRS OEF与侵入性OEF强相关(R=0.61,P=.01;偏差0.09)。
关键数值与推荐
- FD-NIRS StO2与侵入性SvO2的相关系数R=0.74(P<.001),偏差-2.8%。
- CW-NIRS StO2与侵入性SvO2的相关系数R=0.29(P=.34),偏差-2.5%。
- FD-NIRS SaO2与侵入性SaO2的相关系数R=0.84(P<.001),偏差-3.2%。
- 脉搏血氧仪SaO2与侵入性SaO2的相关系数R=0.68(P=.001),偏差0.2%。
- FD-NIRS OEF与侵入性OEF的相关系数R=0.61(P=.01),偏差0.09。
临床怎么用
- 在单心室先天性心脏病儿童中,FD-NIRS可能提供比CW-NIRS和脉搏血氧仪更准确的脑组织氧合和动脉饱和度无创评估。
- FD-NIRS可独立量化OEF,有助于无创评估脑氧摄取情况。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To compare the accuracy of frequency-domain near-infrared spectroscopy (FD-NIRS) with clinically available optical measures of cerebral tissue oxygenation (StO2) and arterial saturation (SaO2). STUDY DESIGN: We enrolled 24 children with single-ventricle congenital heart defects scheduled for cardiac catheterization. Intraoperative FD-NIRS and continuous-wave near-infrared spectroscopy (CW-NIRS) measurements of StO2 were bilaterally obtained on the forehead. SaO2 was simultaneously acquired with FD-NIRS and pulse oximetry. Noninvasive StO2 and SaO2 were compared with invasive superior vena cava (SVC) and descending aorta saturation measurements. These measurements were also used to compute cerebral tissue oxygen extraction fraction (OEF). RESULTS: The FD-NIRS StO2 was strongly correlated with invasive venous oxygen saturation (SvO2) (R = 0.74, P < .001; bias, -2.8%). However, correlation between CW-NIRS StO2 and invasive SvO2 was weaker and not significant (R = 0.29, P = .34; bias, -2.5%). Invasive SaO2 was correlated with FD-NIRS (R = 0.84, P < .001; bias, -3.2%) and pulse oximetry (R = 0.68, P = .001; bias, 0.2%) metrics of SaO2, but pulse oximetry showed larger error at low saturations. Finally, FD-NIRS OEF and invasive OEF were strongly correlated (R = 0.61, P = .01; bias, 0.09). CONCLUSIONS: Our study suggests that FD-NIRS provides more accurate quantification of StO2 and SaO2 than CW-NIRS and pulse oximetry in patients with single-ventricle heart defects. We also demonstrated and validated stand-alone OEF quantification using FD-NIRS.
引用
- Zhao H, Davis C, Anderson D, et al. Non-Invasive Optical Quantification of Cerebral Tissue Oxygenation in Children with Single-Ventricle Congenital Heart Disease. J Pediatr. 2026. PMID: 42476299. doi:10.1016/j.jpeds.2026.115247