新生儿脉搏血氧饱和度准确性与皮肤色素差异:NeoPODS前瞻性研究

Neonatal Pulse Oximetry Accuracy and Disparities by Skin Pigmentation (NeoPODS): A Prospective Study.
PubMed 文献 · J Pediatr · 2026-04-21

这项前瞻性诊断准确性研究在2个三级新生儿重症监护病房纳入70名新生儿,分析136对SpO2-SaO2匹配数据,评估皮肤色素对脉搏血氧饱和度偏差的影响。

要点

  • 分类皮肤色素指标和种族间偏差无显著差异。
  • 连续分析中仅ITA在首次测量时与偏差负相关。

核心结论

  • SpO2轻微低估SaO2但未显示皮肤色素间临床显著差异。

研究设计与样本

  • 前瞻性诊断准确性研究,在2个三级NICU于2022-2025年纳入70名小于10日龄的新生儿,分析136对SpO2-SaO2匹配数据。

主要发现

  • 按分类皮肤色素指标和种族分析,SpO2-SaO2偏差无统计学显著差异。
  • 潜在低氧血症罕见,仅1名浅色皮肤新生儿发生。

临床怎么用

  • 新生儿SpO2读数可能轻微低估SaO2,临床解读时应考虑此偏差。
  • 不同皮肤色素新生儿中SpO2偏差无临床显著差异,支持当前使用。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To assess pulse oximetry saturation (SpO2) to arterial oxygen saturation (SaO2) bias relative to skin pigmentation in newborns and to address potentially occult hypoxemia, defined as SaO2 lower than SpO2. STUDY DESIGN: We conducted a prospective, diagnostic accuracy study at 2 tertiary neonatal intensive care units between 2022-2025 to assess SpO2-SaO2 differences in relation to skin pigmentation in newborns. Newborns younger than 10 postnatal days with an arterial catheter undergoing an arterial blood gas collection were enrolled. A pulse oximeter was placed on the extremity corresponding to the arterial sampling site, and SpO2 values were recorded during blood gas collection. Skin pigmentation was measured noninvasively using individual typology angle (ITA), melanin index, and visual assessments. Mean SpO2 was calculated from the 30 seconds immediately preceding arterial sampling or from the most stable 30-second average when timestamps were unavailable and compared with SaO2 to calculate SpO2-SaO2 bias. RESULTS: A total of 136 matched SpO2-SaO2 pairs from 70 patients were analyzed, of whom 40% were Black. The overall mean bias was -0·98 ± 2·8% (95% CI -1·4 to -0·52), indicating an underestimation of SaO2 by SpO2. Across categorical skin-pigmentation measures and race, no statistically significant differences in SpO2-SaO2 bias were observed. When analyzed continuously, bias became less negative with lighter skin pigmentation for both ITA and melanin index, reaching significance only for ITA when analyses were restricted to the first measurement per patient (slope +0·016 per ITA unit, 95% CI 0·001-0·032, p=0·041). Occult hypoxemia was rare, occurring in one newborn with light skin pigmentation. CONCLUSION: In this prospective neonatal study with tightly paired measurements, SpO2 slightly underestimated SaO2 but did not demonstrate clinically significant disparities across skin pigmentation.

来源:J Pediatr,查看原文。版权归原作者所有。