新生儿低血糖床旁血糖检测方法的系统评价与诊断准确性Meta分析

Glucose Testing Methods: A Systematic Review and Meta-Analysis of Diagnostic Accuracy of Point-of-Care Devices for Neonatal Hypoglycemia.
PubMed 文献 · J Pediatr · 2024-12-13

定量分析(n=31)显示所有方法特异性≥93%,葡萄糖氧化酶+电化学、葡萄糖-1-脱氢酶+电化学和己糖激酶+电化学敏感性更优,汇总ROC曲线确认己糖激酶+电化学最准确。

要点

  • 所有床旁血糖检测方法特异性均≥93%,适用于排除低血糖。
  • 葡萄糖氧化酶+电化学、葡萄糖-1-脱氢酶+电化学和己糖激酶+电化学敏感性更优。
  • 己糖激酶+电化学在汇总ROC曲线中显示最准确,但研究数量有限。

核心结论

  • 某些床旁血糖检测方法在测量新生儿血糖浓度时具有更高的准确性,其中己糖激酶+电化学方法最可靠。

主要发现

  • 定量分析评估了葡萄糖氧化酶+光度法(n=8)、葡萄糖-1-脱氢酶+光度法(n=6)、葡萄糖氧化酶+电化学(n=13)、葡萄糖-1-脱氢酶+电化学(n=12)和己糖激酶+电化学(n=2)。
  • 所有方法显示高特异性(≥93%),葡萄糖氧化酶+电化学、葡萄糖-1-脱氢酶+电化学和己糖激酶+电化学显示更优的敏感性。
  • 汇总受试者工作特征曲线确认己糖激酶+电化学最准确。

关键数值与推荐

  • 所有方法特异性≥93%。
  • 葡萄糖氧化酶+电化学、葡萄糖-1-脱氢酶+电化学和己糖激酶+电化学敏感性更优。
  • 己糖激酶+电化学在汇总ROC曲线中显示最准确。

临床怎么用

  • 临床选择床旁血糖检测方法时可优先考虑己糖激酶+电化学方法以提高准确性。
  • 所有方法高特异性可用于排除低血糖,但敏感性差异影响低血糖检出。

局限性

  • 己糖激酶+电化学方法的研究数量有限,需进一步研究在不同环境和人群中确认。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To evaluate the accuracy of various point-of-care device methodologies for measuring blood glucose concentrations in babies at risk of neonatal hypoglycemia. STUDY DESIGN: This systematic review and meta-analysis included studies from Ovid MEDLINE, Ovid Embase, and Web of Science up to May 20, 2024. Studies comparing point-of-care testing methods for neonatal blood glucose to a standard laboratory method were included, excluding those on continuous glucose monitoring or conducted before 1990. Two researchers independently assessed inclusion and evaluated risk of bias using QUADAS-2. Sensitivity and specificity were calculated using contingency tables, and diagnostic accuracy was analyzed using hierarchical random-effects modelling. Studies with insufficient data were summarized by estimation direction. RESULTS: Seventy-one studies were included. The quantitative analysis (n = 31) evaluated glucose oxidase (GO) + photometry (n = 8), glucose-1-dehydrogenase (GDH) + photometry (n = 6), GO + electrochemistry (n = 13), GDH + electrochemistry (n = 12), and hexokinase (HK) + electrochemistry (n = 2). All methods showed high specificity (≥93%), with GO + electrochemistry, GDH + electrochemistry, and HK + electrochemistry showing superior sensitivity. The summary receiver operating characteristic curve confirmed HK + electrochemistry as the most accurate. CONCLUSION: Certain point-of-care device methodologies demonstrate greater accuracy in measuring neonatal blood glucose concentrations. Of the methods evaluated, HK + electrochemistry proved to be the most reliable. However, the limited number of studies using this method suggests the need for further research to confirm these findings across diverse settings and populations. SYSTEMATIC REVIEW REGISTRATION: PROSPERO on December 29, 2023 (CRD42023488539).

来源:J Pediatr,查看原文。版权归原作者所有。

引用

  1. St Clair SL, Ulyatt CM, Corkin MT, et al. Glucose Testing Methods: A Systematic Review and Meta-Analysis of Diagnostic Accuracy of Point-of-Care Devices for Neonatal Hypoglycemia. J Pediatr. 2024. PMID: 39675663. doi:10.1016/j.jpeds.2024.114438

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。