文献速递心血管DSA

基于分叉分形定律的冠脉造影衍生FFR对冠脉狭窄血流动力学意义的诊断效能

Diagnostic performance of angiography-derived fractional flow reserve analysis based on bifurcation fractal law for assessing hemodynamic significance of coronary stenosis.
PubMed 文献 · Eur Radiol · 2023-05-03

本研究提出一种基于分叉分形定律补偿边支血流的冠脉造影衍生FFR(Angio-FFR)分析方法,在119例患者159条血管中与考虑边支的FFR_s和忽略边支的FFR_n比较,并以有创FFR为参考。结果显示Angio-FFR诊断准确性与FFR_s相当,显著优于FFR_n,与有创FFR相关性良好。

要点

  • Angio-FFR通过分叉分形定律补偿边支血流,无需勾画边支即可评估主支狭窄功能意义。
  • 其诊断准确性与勾画边支的FFR_s相当,显著优于忽略边支的FFR_n。
  • 与有创FFR相关性良好(r=0.92),支持其用于血流动力学评估。

核心结论

  • 基于分叉分形定律补偿边支血流的Angio-FFR在评估冠脉狭窄血流动力学意义方面具有良好的诊断性能。

研究设计与样本

  • 单中心回顾性研究,纳入119例患者共159条血管,采用基于血管段的一维降阶模型进行Angio-FFR分析。
  • 以有创FFR为参考标准,比较Angio-FFR与FFR_s(勾画冠脉树考虑边支血流)及FFR_n(仅勾画主支忽略边支血流)的诊断准确性。

诊断标准与关键数值

  • Angio-FFR与有创FFR的Pearson相关系数为0.92,FFR_s为0.91,FFR_n仅为0.85。
  • Angio-FFR诊断准确性与FFR_s相当,且显著高于FFR_n。

临床应用与阅片要点

  • 在Angio-FFR计算中可利用分叉分形定律补偿边支血流,无需勾画边支即可评估主支狭窄功能严重程度。
  • 忽略边支血流可能降低Angio-FFR的诊断准确性,阅片时应注意边支血流的影响。

局限性

  • 原文未明确报告样本量估算、多中心验证及具体敏感度、特异度、AUC等诊断效能指标。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Blood flow into the side branch affects the calculation of coronary angiography-derived fractional flow reserve (FFR), called Angio-FFR. Neglecting or improperly compensating for the side branch flow may decrease the diagnostic accuracy of Angio-FFR. This study aims to evaluate the diagnostic accuracy of a novel Angio-FFR analysis that considers the side branch flow based on the bifurcation fractal law.

METHODS: A one-dimensional reduced-order model based on the vessel segment was used to perform Angio-FFR analysis. The main epicardial coronary artery was divided into several segments according to the bifurcation nodes. Side branch flow was quantified using the bifurcation fractal law to correct the blood flow in each vessel segment. In order to verify the diagnostic performance of our Angio-FFR analysis, two other computational methods were taken as control groups: (i) FFR_s: FFR calculated by delineating the coronary artery tree to consider side branch flow, (ii) FFR_n: FFR calculated by just delineating the main epicardial coronary artery and neglecting the side branch flow.

RESULTS: The analysis of 159 vessels from 119 patients showed that our Anio-FFR calculation method had comparable diagnostic accuracy to FFR_s and provided significantly higher diagnostic accuracy than that of FFR_n. In addition, using invasive FFR as a reference, the Pearson correlation coefficients of Angio-FFR and FFRs were 0.92 and 0.91, respectively, while that of FFR_n was only 0.85.

CONCLUSIONS: Our Angio-FFR analysis has demonstrated good diagnostic performance in assessing the hemodynamic significance of coronary stenosis by using the bifurcation fractal law to compensate for side branch flow.

CLINICAL RELEVANCE STATEMENT: Bifurcation fractal law can be used to compensate for side branch flow during the Angio-FFR calculation of the main epicardial vessel. Compensating for side branch flow can improve the ability of Angio-FFR to diagnose stenosis functional severity.

KEY POINTS: • The bifurcation fractal law could accurately estimate the blood flow from the proximal main vessel into the main branch, thus compensating for the side branch flow. • Angiography-derived FFR based on the bifurcation fractal law is feasible to evaluate the target diseased coronary artery without delineating the side branch.

来源:Eur Radiol,查看原文。版权归原作者所有。