文献速递心血管CTDSA

基于深度学习的高速现场FFR-CT算法:以有创血管造影为参考标准的评估

High-Speed On-Site Deep Learning-Based FFR-CT Algorithm: Evaluation Using Invasive Angiography as the Reference Standard.
PubMed 文献 · AJR Am J Roentgenol · 2023-05-03

本回顾性研究纳入59例患者74个病变,评估现场深度学习FFR-CT算法诊断血流动力学显著狭窄的性能,以有创FFR≤0.80和/或瞬时无波形比值≤0.89为参考标准。结果显示AUC达0.975,阈值0.80时准确度95.9%、敏感度93.5%、特异度97.7%,平均分析时间7分54秒,观察者间一致性良好至优秀。

要点

  • 现场深度学习FFR-CT算法诊断血流动力学显著狭窄的AUC为0.975,准确度95.9%。
  • 严重钙化(≥400 Agatston单位)病变中AUC仍达0.991,敏感度94.7%、特异度95.0%。
  • 平均每例分析时间7分54秒,观察者内和观察者间一致性良好至优秀。

局限性

  • 回顾性研究设计。
  • 单中心、样本量较小(59例患者74个病变)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND. Estimation of fractional flow reserve from coronary CTA (FFR-CT) is an established method of assessing the hemodynamic significance of coronary lesions. However, clinical implementation has progressed slowly, partly because of off-site data transfer with long turnaround times for results. OBJECTIVE. The purpose of this study was to evaluate the diagnostic performance of FFR-CT computed on-site with a high-speed deep learning-based algorithm with invasive hemodynamic indexes as the reference standard. METHODS. This retrospective study included 59 patients (46 men, 13 women; mean age, 66.5 ± 10.2 years) who underwent coronary CTA (including calcium scoring) followed within 90 days by invasive angiography with invasive fractional flow reserve (FFR) and/or instantaneous wave-free ratio measurements from December 2014 to October 2021. Coronary artery lesions were considered to have hemodynamically significant stenosis in the presence of invasive FFR of 0.80 or less and/or instantaneous wave-free ratio of 0.89 or less. A single cardiologist evaluated the CTA images using an on-site deep learning-based semiautomated algorithm entailing a 3D computational flow dynamics model to determine FFR-CT for coronary artery lesions detected with invasive angiography. Time for FFR-CT analysis was recorded. FFR-CT analysis was repeated by the same cardiologist in 26 randomly selected examinations and by a different cardiologist in 45 randomly selected examinations. Diagnostic performance and agreement were assessed. RESULTS. A total of 74 lesions were identified with invasive angiography. FFR-CT and invasive FFR had strong correlation (r = 0.81) and, in Bland-Altman analysis, bias of 0.01 and 95% limits of agreement of -0.13 to 0.15. FFR-CT had AUC for hemodynamically significant stenosis of 0.975. At a cutoff of 0.80 or less, FFR-CT had 95.9% accuracy, 93.5% sensitivity, and 97.7% specificity. In 39 lesions with severe calcifications (≥ 400 Agatston units), FFR-CT had AUC of 0.991 and at a cutoff of 0.80, 94.7% sensitivity, 95.0% specificity, and 94.9% accuracy. Mean analysis time per patient was 7 minutes 54 seconds. Intraobserver agreement (intraclass correlation coefficient, 0.85; bias, -0.01; 95% limits of agreement, -0.12 and 0.10) and interobserver agreement (intraclass correlation coefficient, 0.94; bias, -0.01; 95% limits of agreement, -0.08 and 0.07) were good to excellent. CONCLUSION. A high-speed on-site deep learning-based FFR-CT algorithm had excellent diagnostic performance for hemodynamically significant stenosis with high reproducibility. CLINICAL IMPACT. The algorithm should facilitate implementation of FFR-CT technology into routine clinical practice.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。

引用

  1. Giannopoulos, Keller, Sepulcri, et al. High-Speed On-Site Deep Learning-Based FFR-CT Algorithm: Evaluation Using Invasive Angiography as the Reference Standard. AJR Am J Roentgenol. 2023. PMID: 37132550. doi:10.2214/AJR.23.29156

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。