磁共振引导高强度聚焦超声评估声敏全氟化碳乳剂在灌流组织仿体中的声吸收增强
本研究在湍流灌流组织仿体中循环不同浓度的F8TAC18-PFOB声敏乳剂,用磁共振测温监测聚焦超声加热,并以生物热方程模拟量化声吸收。结果显示0.25% v:v浓度下声吸收系数从0.93 ± 0.05增至1.82 ± 0.22 m-1(0.1 mL s-1流速),无灌注时从1.23 ± 0.07增至1.65 ± 0.17 m-1,进一步支持该乳剂的热增强效应。
要点
- 该乳剂在湍流灌流仿体中可浓度依赖性增加声吸收系数。
- 0.25% v:v浓度下声吸收系数较对照增高约一倍(0.1 mL s-1流速)。
- 磁共振测温联合生物热方程模拟可用于预测未测试条件下的热效应。
核心结论
- F8TAC18-PFOB声敏乳剂在湍流灌流仿体中可增强聚焦超声的声吸收,且该效应可在多种内脏组织模型中被观察到。
研究设计与样本
- 研究类型为体外组织仿体实验,使用湍流灌流仿体循环F8TAC18-PFOB乳剂,浓度梯度为0.07、0.13、0.19和0.25% v:v。
- 采用磁共振测温监测聚焦超声(67 W、95%占空比、33 s)加热过程,并以生物热方程进行在体模拟。
- 对照条件为循环去气去离子水(无液滴),乳剂液滴直径为1.24 µm。
诊断标准与关键数值
- 在0.1 mL s-1流速下,声吸收系数从0.0% v:v的0.93 ± 0.05 m-1增至0.25% v:v的1.82 ± 0.22 m-1。
- 无灌注条件下,0.25% v:v的声吸收系数从0.0% v:v的1.23 ± 0.07 m-1增至1.65 ± 0.17 m-1。
临床应用与阅片要点
- 该乳剂可能作为聚焦超声热疗的增敏剂,阅片时需关注磁共振测温所显示的温升变化。
- 声吸收系数的量化方法可用于模拟不同灌注条件下的热效应,辅助治疗计划。
局限性
- 研究为体外仿体实验,结果不能直接外推至在体组织。
- 仅测试了有限浓度范围和单一流速条件。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Sonosensitive high-boiling point perfluorocarbon F8TAC18-PFOB emulsions previously exhibited thermal enhancement during focused ultrasound heating in ex vivo pig livers, kidneys and a laminar flow phantom. The main objectives of this study were to evaluate heating under turbulent conditions, observe perfusion effects, quantify heating in terms of acoustic absorption and model the experimental data.
METHODS: In this study, similar perfluorocarbon emulsions were circulated at incremental concentrations of 0.07, 0.13, 0.19 and 0.25% v:v through a percolated turbulent flow phantom, more representative of the biological tissue than a laminar flow phantom. The concentrations represent the droplet content in only the perfused fluid, rather than the droplet concentration throughout the entire cross-section. The temperature was measured with magnetic resonance thermometry, during focused ultrasound sonications of 67 W, 95% duty cycle and 33 s duration. These were used in Bioheat equation simulations to investigate in silico the thermal phenomena. The temperature change was compared with the control condition by circulating de-gassed and de-ionized water through the flow phantom without droplets.
RESULTS: With these 1.24 µm diameter droplets at 0.25% v:v, the acoustic absorption coefficient increased from 0.93 ± 0.05 at 0.0% v:v to 1.82 ± 0.22 m-1 at 0.25% v:v using a 0.1 mL s-1 flow rate. Without perfusion at 0.25% v:v, an increase was observed from 1.23 ± 0.07 m-1 at 0.0% v:v to 1.65 ± 0.17 m-1.
CONCLUSION: The results further support previously reported thermal enhancement with F8TAC18-PFOB emulsion, quantified the increased absorption at small concentration intervals, illustrated that the effects can be observed in a variety of visceral tissue models and provided a method to simulate untested scenarios.
引用
- Holman, Guillemin, Lorton, et al. Assessing Enhanced Acoustic Absorption From Sonosensitive Perfluorocarbon Emulsion With Magnetic Resonance-Guided High-Intensity Focused Ultrasound and a Percolated Tissue-Mimicking Flow Phantom. Ultrasound Med Biol. 2023. PMID: 37117139. doi:10.1016/j.ultrasmedbio.2023.01.022