儿童高分辨率胸部CT:光子计数探测器与能量积分探测器辐射剂量及图像质量匹配研究

Comparison of Radiation Dose and Image Quality of Pediatric High-Resolution Chest CT Between Photon-Counting Detector CT and Energy-Integrated Detector CT: A Matched Study.
PubMed 文献 · AJR Am J Roentgenol · 2023-04-05

PCD CT的CTDIvol、DLP、SSDE及管电流-曝光时间乘积均显著低于EID CT,而客观图像质量(肺衰减、噪声、SNR)和主观图像质量(整体质量、运动伪影)无显著差异。结论支持PCD CT在儿童中常规使用。

要点

  • 儿童胸部HRCT中,PCD CT较EID CT显著降低辐射剂量。
  • 两组客观图像质量(肺衰减、噪声、SNR)无显著差异。
  • 两组主观图像质量及运动伪影评分无显著差异。

局限性

  • 回顾性研究设计。
  • 单中心小样本(每组27例)。
  • 仅纳入儿童胸部HRCT,结论外推需谨慎。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND. Photon-counting detector (PCD) CT has been shown to reduce radiation dose and improve image quality in adult chest CT examinations; its potential impact in pediatric CT is not well documented. OBJECTIVE. The purpose of our study was to compare radiation dose, objective image quality, and subjective image quality of PCD CT and energy-integrating detector (EID) CT in children undergoing high-resolution CT (HRCT) of the chest. METHODS. This retrospective study included 27 children (median age, 3.9 years; 10 girls, 17 boys) who underwent PCD CT between March 1, 2022, and August 31, 2022, and 27 children (median age, 4.0 years; 13 girls, 14 boys) who underwent EID CT between August 1, 2021, and January 31, 2022; all examinations comprised clinically indicated chest HRCT. The patients in the two groups were matched by age and water-equivalent diameter. Radiation dose parameters were recorded. One observer placed ROIs to measure objective parameters (lung attenuation, image noise, and SNR). Two radiologists independently assessed subjective measures (overall image quality and motion artifacts) using 5-point Likert scales (1 = highest quality). Groups were compared. RESULTS. PCD CT, in comparison with EID CT, showed lower median CTDIvol (0.41 vs 0.71 mGy, p < .001), DLP (10.2 vs 13.7 mGy × cm, p = .008), size-specific dose estimate (0.82 vs 1.34 mGy, p < .001), and tube current-exposure time product (48.0 vs 202.0 mAs, p < .001). PCD CT and EID CT showed no significant difference in right upper lobe (RUL) lung attenuation (mean, -793 vs -750 HU; p = .09), right lower lobe (RLL) lung attenuation (mean, -745 vs -716 HU; p = .23), RUL image noise (mean, 55 vs 51 HU; p = .27), RLL image noise (mean, 59 vs 57 HU; p = .48), RUL SNR (mean, -14.9 vs -15.8; p = .89), or RLL SNR (mean, -13.1 vs -13.6; p = .79). PCD CT and EID CT showed no significant difference in median overall image quality for reader 1 (1.0 vs 1.0, p = .28) or reader 2 (1.0 vs 1.0, p = .17) or median motion artifacts for reader 1 (1.0 vs 1.0, p = .07) or reader 2 (1.0 vs 1.0, p = .22). CONCLUSION. PCD CT showed significantly reduced dose levels without a significant difference in objective or subjective image quality compared with EID CT. CLINICAL IMPACT. These data expand understanding of the capabilities of PCD CT and support its routine use in children.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。

引用

  1. Siegel, Bugenhagen, Sanchez, et al. Comparison of Radiation Dose and Image Quality of Pediatric High-Resolution Chest CT Between Photon-Counting Detector CT and Energy-Integrated Detector CT: A Matched Study. AJR Am J Roentgenol. 2023. PMID: 37095666. doi:10.2214/AJR.23.29077

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。