经鼻肠梗阻导管置入术辐射剂量及超诊断参考水平相关因素:REX-GI研究496例分析

Radiation dose and factors related to exceeding the diagnostic reference level in 496 transnasal ileus tube placement procedures from the REX-GI study.
PubMed 文献 · Br J Radiol · 2023-04-22

基于REX-GI研究496例经鼻肠梗阻导管置入术患者,计算PKA第三四分位值47 Gycm2作为日本诊断参考水平(DRL)。超DRL者透视时间、操作时间和导管插入长度显著更长,提示较长操作时间和插入长度与超DRL相关。

要点

  • 经鼻肠梗阻导管置入术的DRL设为PKA 47 Gycm2(第三四分位值)。
  • 超DRL者透视时间19 vs 10 min、操作时间25 vs 18 min、插入长度185 vs 165 cm。
  • 操作时间延长和导管插入长度增加与辐射剂量超DRL相关。

核心结论

  • 经鼻肠梗阻导管置入术的DRL为PKA 47 Gycm2,较长操作时间和导管插入长度可能与超DRL相关。

研究设计与样本

  • 记录内镜插入时间、操作时间、导管插入长度、透视时间、Ka.r和PKA,以PKA第三四分位值作为DRL。
  • 分析超DRL的相关因素,并比较超DRL与未超DRL组的参数差异。

诊断标准与关键数值

  • DRL定义为PKA第三四分位值47 Gycm2;超DRL组透视时间19 min vs 10 min、操作时间25 min vs 18 min、导管插入长度185 cm vs 165 cm,差异显著。
  • 全体患者FT第三四分位/中位数为18/11.9 min,Ka.r为99.2/54.4 mGy,PKA为46.9/28 Gycm2。
  • 内镜插入时间、操作时间和导管插入长度中位数分别为6 min、20 min和170 cm。

临床应用与阅片要点

  • 经鼻肠梗阻导管置入术中应关注透视时间和PKA,操作时间延长或导管插入长度增加时警惕辐射剂量超DRL。
  • 不同医院间辐射剂量分布存在差异,提示需结合本机构数据优化防护和操作流程。

局限性

  • 本文为REX-GI研究的事后分析,原文未报告多中心设计细节以外的其他局限。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: We aimed to examine the factors contributing to radiation exposure exceeding the DRL of the transnasal ileus tube placement in this post hoc analysis from the cohort of the REX-GI study.

METHODS: Patients with transnasal ileus tubes were enrolled in the rex-gi study from may 2019 to december 2020. We investigated the endoscope insertion time (min), procedure time (min), tube insertion length (cm), fluoroscopy time (FT: min), air kerma at the patient entrance reference point (Ka.r: mGy), and air kerma-area product (PKA: Gycm2). The third quartile value of the PKA value was calculated as the diagnostic reference level (DRL) value. We explored the factors associated with radiation exposure exceeding the DRL.

RESULTS: In the REX-GI study, 496 patients who underwent transnasal ileus tube placement were enrolled. The median age of the patients was 71 years. The median endoscopy insertion time, procedure time, and tube insertion length were 6 min, 20 min, and 170 cm, respectively. The third quartile/median FT, Ka.r, and PKA were 18/11.9 min, 99.2/54.4 mGy, and 46.9/28 Gycm2, respectively. The third quartile value of PKA (47 Gycm2) was set as the DRL value. There were differences in distribution by the hospital. Compared with procedures under the DRL, the FT (19 vs 10 min), procedure time (25 vs 18 min), and tube insertion length (185 vs 165 cm) were significantly longer for procedures above the DRL.

CONCLUSION: We report the DRL for transnasal ileus tube placement in Japan. A longer procedure time and tube insertion length may be associated with DRL exceedance.

ADVANCES IN KNOWLEDGE: Transnasal ileus tube placement under fluoroscopy guidance is a standard clinical procedure for bowel obstruction. However, the appropriate radiation dose level has not yet been established.We report the (DRL) for transnasal ileus tube placement in Japan. A longer procedure time and tube insertion length may be associated with DRL exceedance.

来源:Br J Radiol,查看原文。版权归原作者所有。