质子治疗中基于CT的停止本领比预测:Hounsfield查找表共识指南
该共识指南提出六步法标准化Hounsfield查找表(HLUT)的制定流程,包括体模设置、CT采集、CT值提取、SPR测定、HLUT制定与验证。指南强调结合体模插入物与表格化人体组织的SPR可提高HLUT稳定性,并在三家不同厂商CT扫描仪上验证了可行性,旨在减少欧洲质子中心间SPR预测的差异。
要点
- HLUT制定需六步:体模设置、CT采集、CT值提取、SPR测定、HLUT制定与验证。
- CT值提取应涵盖每个插入物平面内内部70%的感兴趣区及扫描方向多个轴位层面。
- 结合体模插入物与100 MeV化学计量法计算的人体组织SPR,可提高HLUT稳定性。
诊断标准与关键数值
- CT值提取自覆盖每个插入物平面内内部70%的感兴趣区及扫描方向多个轴位CT层面。
- 体模插入物的SPR在质子束中测量,表格化人体组织的SPR在100 MeV下化学计量法计算。
- 对肺、脂肪、软组织、骨四个组织组分别进行CT值与SPR的分段线性回归,再用直线连接。
临床应用与阅片要点
- 日常应按照六步流程规范制定和验证HLUT,以提高SPR预测一致性。
- 制定HLUT时同时纳入体模插入物和表格化人体组织数据,以增强稳定性。
- 验证步骤应简单但全面,确保HLUT在临床扫描仪上可行。
局限性
- 指南仅在三家CT扫描仪上示例测试,未报告大样本多中心验证结果。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Studies have shown large variations in stopping-power ratio (SPR) prediction from computed tomography (CT) across European proton centres. To standardise this process, a step-by-step guide on specifying a Hounsfield look-up table (HLUT) is presented here.
MATERIALS AND METHODS: The HLUT specification process is divided into six steps: Phantom setup, CT acquisition, CT number extraction, SPR determination, HLUT specification, and HLUT validation. Appropriate CT phantoms have a head- and body-sized part, with tissue-equivalent inserts in regard to X-ray and proton interactions. CT numbers are extracted from a region-of-interest covering the inner 70% of each insert in-plane and several axial CT slices in scan direction. For optimal HLUT specification, the SPR of phantom inserts is measured in a proton beam and the SPR of tabulated human tissues is computed stoichiometrically at 100 MeV. Including both phantom inserts and tabulated human tissues increases HLUT stability. Piecewise linear regressions are performed between CT numbers and SPRs for four tissue groups (lung, adipose, soft tissue, and bone) and then connected with straight lines. Finally, a thorough but simple validation is performed.
RESULTS: The best practices and individual challenges are explained comprehensively for each step. A well-defined strategy for specifying the connection points between the individual line segments of the HLUT is presented. The guide was tested exemplarily on three CT scanners from different vendors, proving its feasibility.
CONCLUSION: The presented step-by-step guide for CT-based HLUT specification with recommendations and examples can contribute to reduce inter-centre variations in SPR prediction.
引用
- Peters, Trier Taasti, Ackermann, et al. Consensus guide on CT-based prediction of stopping-power ratio using a Hounsfield look-up table for proton therapy. Radiother Oncol. 2023. PMID: 37084884. doi:10.1016/j.radonc.2023.109675