新型降噪算法对肥胖患者双能腹部CT图像质量的影响评估
测试算法较标准算法和对比算法获得更高的客观图像质量(CNR),且阅片者主观偏好测试算法,但在图像质量和诊断舒适度评分上各算法间无统计学显著差异。
要点
- 测试降噪算法可显著提高肥胖患者双能腹部CT的客观图像质量(CNR),但主观图像质量与诊断舒适度提升有限。
- 诊断准确率在各算法间均较低(66%、63%、67%),提示降噪不改变诊断准确性。
- 阅片者对测试算法降噪后的图像噪声评分明显优于标准算法和对比算法。
关键影像表现/征象
- 平均CNR:原始图像5.0±1.9,对比算法31.1±10.3(p<0.001),测试算法8.9±2.9(p<0.001)。
- 阅片者间一致性:感知图像噪声ICC为0.83(良好),图像质量ICC为0.71(中等至良好),诊断舒适度ICC为0.6(中等)。
- 主观评分均值:噪声评分原始54、对比84、测试66(p<0.05);图像质量评分原始59、对比61、测试65;诊断舒适度评分原始63、对比68、测试68。图像质量和诊断舒适度评分在算法间无统计学显著差异。
诊断标准与关键数值
- 诊断准确率:原始算法66%,对比算法63%,测试算法67%(原文未报告各算法间差异的统计学检验结果)。
- 主观评分标准:0分(最差)至100分(最佳),评估感知图像噪声、图像质量和诊断舒适度。
局限性
- 回顾性研究设计。
- 诊断准确率整体较低(63%-67%),可能限制对算法临床价值的判断。
- 主观图像质量和诊断舒适度在算法间无显著差异,降噪的临床获益需进一步验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Evaluate a novel algorithm for noise reduction in obese patients using dual-source dual-energy (DE) CT imaging.
METHODS: Seventy-nine patients with contrast-enhanced abdominal imaging (54 women; age: 58 ± 14 years; BMI: 39 ± 5 kg/m2, range: 35-62 kg/m2) from seven DECT (SOMATOM Flash or Force) were retrospectively included (01/2019-12/2020). Image domain data were reconstructed with the standard clinical algorithm (ADMIRE/SAFIRE 2), and denoised with a comparison (ME-NLM) and a test algorithm (rank-sparse kernel regression). Contrast-to-noise ratio (CNR) was calculated. Four blinded readers evaluated the same original and denoised images (0 (worst)-100 (best)) in randomized order for perceived image noise, quality, and their comfort making a diagnosis from a table of 80 options. Comparisons between algorithms were performed using paired t-tests and mixed-effects linear modeling.
RESULTS: Average CNR was 5.0 ± 1.9 (original), 31.1 ± 10.3 (comparison; p < 0.001), and 8.9 ± 2.9 (test; p < 0.001). Readers were in good to moderate agreement over perceived image noise (ICC: 0.83), image quality (ICC: 0.71), and diagnostic comfort (ICC: 0.6). Diagnostic accuracy was low across algorithms (accuracy: 66, 63, and 67% (original, comparison, test)). The noise received a mean score of 54, 84, and 66 (p < 0.05); image quality 59, 61, and 65; and the diagnostic comfort 63, 68, and 68, respectively. Quality and comfort scores were not statistically significantly different between algorithms.
CONCLUSIONS: The test algorithm produces quantitatively higher image quality than current standard and existing denoising algorithms in obese patients imaged with DECT and readers show a preference for it.
CLINICAL RELEVANCE STATEMENT: Accurate diagnosis on CT imaging of obese patients is challenging and denoising algorithms can increase the diagnostic comfort and quantitative image quality. This could lead to better clinical reads.
KEY POINTS: • Improving image quality in DECT imaging of obese patients is important for accurate and confident clinical reads, which may be aided by novel denoising algorithms using image domain data. • Accurate diagnosis on CT imaging of obese patients is especially challenging and denoising algorithms can increase quantitative and qualitative image quality. • Image domain algorithms can generalize well and can be implemented at other institutions.