Nakagami成像定量分析用于高强度聚焦超声治疗病灶的精准检测与监测建模
该研究通过仿真和体模、离体猪肝实验,探讨Nakagami成像检测高强度聚焦超声热消融病灶的准确性。仿真中病灶横向和轴向估计尺寸分别为4.91和4.79 mm,接近实际5×5 mm;亚分辨效应主要发生于气泡与未治疗区边界,不影响检测准确性。实验证实Nakagami成像能准确区分热病灶与气泡,精度取决于边界划分所选Nakagami值。
要点
- Nakagami成像可准确识别热消融病灶并区分气泡。
- 亚分辨效应有助于病灶识别,精度取决于边界Nakagami阈值选择。
- 目前证据来自仿真和离体实验,尚需在体及临床验证。
核心结论
- Nakagami成像能准确检测高强度聚焦超声热消融病灶,亚分辨效应有助于病灶识别。
研究设计与样本
- 研究包含仿真、体模实验和离体猪肝热消融实验三部分。
- 仿真中设定热病灶实际尺寸为5×5 mm,用于评估Nakagami成像的检测精度。
关键影像表现/征象
- 仿真中Nakagami图像估计的病灶横向和轴向尺寸分别为4.91和4.79 mm,接近实际5×5 mm。
- 亚分辨区域主要出现在气泡与未治疗区之间的边界,不影响热病灶检测的准确性。
- 热消融实验证实Nakagami成像能够从气泡中准确识别热病灶。
诊断标准与关键数值
- 热病灶检测的精度与Nakagami成像中用于边界划分的Nakagami值选择相关。
临床应用与阅片要点
- 在HIFU治疗监测中可利用Nakagami成像区分热病灶与气泡,但需注意边界划分阈值的选择会影响病灶尺寸估计。
局限性
- 研究仅基于仿真、体模和离体猪肝实验,尚需在体和临床研究进一步验证其应用潜力。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Nakagami imaging is an appealing monitoring and evaluation technique for high-intensity focused ultrasound treatment when bubbles are present in ultrasound images. This study aimed to investigate the accuracy of thermal lesion detection using Nakagami imaging.
METHODS: Simulations were conducted to explore and quantify the influence of the bubbles and the subresolvable effect at the boundary of the thermal lesion on thermal lesion detection. The thermal ablation experiments were conducted in phantom and porcine liver ex vivo.
RESULTS: In the simulation, the estimated lateral and axial size of the thermal lesion in the Nakagami image was 4.91 and 4.79 mm, close to the actual size (5 × 5 mm). The simulation results indicated that the subresolvable region in high-intensity focused ultrasound treatment thermal ablation mainly happened at the boundary between bubbles and the untreated region and does not affect the accuracy of thermal lesion detection. The accurate detection of the thermal lesion using Nakagami imaging mainly depends on bubbles and thermal lesion characterization. Our thermal ablation experiments confirmed that Nakagami imaging has the ability to accurately identify thermal lesions from bubbles.
CONCLUSION: The subresolvable effect is helpful for thermal lesion identification, and precision is related to the Nakagami values chosen for boundary division in Nakagami imaging. Therefore, Nakagami imaging is a promising method for accurately evaluating thermal lesions. Further studies in vivo and in clinical settings will be needed to explore its potential applications.
引用
- Han, Song, Zhang, et al. Modeling for Quantitative Analysis of Nakagami Imaging in Accurate Detection and Monitoring of Therapeutic Lesions by High-Intensity Focused Ultrasound. Ultrasound Med Biol. 2023. PMID: 37080865. doi:10.1016/j.ultrasmedbio.2023.03.002
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。