肺部双能CT:不同技术方案对定量分析的影响

Lung dual energy CT: Impact of different technological solutions on quantitative analysis.
PubMed 文献 · Eur J Radiol · 2023-04-06

该研究采用专用体模评估四种不同双能CT扫描仪(双源、TwinBeam、序列单源)的肺部光谱参数定量准确性。结果显示双源CT和序列单源CT的物质分离角度更大、准确性更好,TwinBeam性能较差;碘浓度测量因肺结节碘含量低而变异性高,临床实用性有限。

要点

  • 双源和序列单源CT的物质分离能力优于TwinBeam。
  • TwinBeam的电子密度、有效原子序数和碘浓度准确性最差。
  • 肺结节碘浓度测量变异性高,临床解读需谨慎。

诊断标准与关键数值

  • 双源CT的钙碘信号角度分离α=9.7°,序列单源CT为α=9.9°,TwinBeam为α=5.0°。
  • TwinBeam准确性最差:Aρe=4.7±1.0%,AZ=9.1±3.1%,AIconc=19.4±4.4%。
  • Flash准确性最佳:Aρe=1.8±0.4%,AZ=3.5±0.7%,AIconc=7.3±1.8%。
  • TwinBeam精密度较差:Pρe=0.6±0.1%,PZ=1.1±0.2%,PIconc=10.9±4.9%,其他技术平均精密度相当。

局限性

  • 研究仅使用体模,未在临床患者中验证。
  • 碘浓度测量因肺结节碘绝对含量低而变异性高,临床实用性受限。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluated the accuracy of spectral parameters quantification of four different CT scanners in dual energy examinations of the lung using a dedicated phantom.

METHOD: Measurements were made with different technologies of the same vendor: one dual source CT scanner (DSCT), one TwinBeam (i.e. split filter) and two sequential acquisition single source scanners (SSCT). Angular separation of Calcium and Iodine signals were calculated from scatter plots of low-kVp versus high-kVp HUs. Electron density (ρe), effective atomic number (Zeff) and Iodine concentration (Iconc) were measured using Syngo.via software. Accuracy (A) of ρe, Zeff and Iconc was evaluated as the absolute percentage difference (D%) between reference values and measured ones, while precision (P) was evaluated as the variability σ obtained by repeating the measurement with different acquisition/reconstruction settings.

RESULTS: Angular separation was significantly larger for DSCT (α = 9.7°) and for sequential SSCT (α = 9.9°) systems. TwinBeam was less performing in material separation (α = 5.0°). The lowest average A was observed for TwinBeam (Aρe = [4.7 ± 1.0], AZ = [9.1 ± 3.1], AIconc = [19.4 ± 4.4]), while the best average A was obtained for Flash (Aρe = [1.8 ± 0.4], AZ = [3.5 ± 0.7], AIconc = [7.3 ± 1.8]). TwinBeam presented inferior average P (Pρe = [0.6 ± 0.1], PZ = [1.1 ± 0.2], PIconc = [10.9 ± 4.9]), while other technologies demonstrate a comparable average.

CONCLUSIONS: Different technologies performed material separation and spectral parameter quantification with different degrees of accuracy and precision. DSCT performed better while TwinBeam demonstrated not excellent performance. Iodine concentration measurements exhibited high variability due to low Iodine absolute content in lung nodules, thus limiting its clinical usefulness in pulmonary applications.

来源:Eur J Radiol,查看原文。版权归原作者所有。