双能CT虚拟单能成像提高EVAR术后内漏检出
双期方案(VNC+动脉期+延迟期)VMI敏感度100%,较参考方案检出更多内漏,平均有效剂量降低23.28%;单期方案敏感度83.33%,剂量降低61.37%。
要点
- 双期VMI方案对EVAR术后内漏检出敏感度达100%,可替代传统三期方案。
- 双期VMI方案平均有效剂量降低23.28%,单期方案降低61.37%但敏感度降至83.33%。
- VNC图像可替代真实平扫,简化协议减少辐射剂量。
核心结论
- 使用VMI的双期简化方案可提高内漏检出数量,敏感度100%,并可能将辐射剂量降低至约四分之一。
- 进一步简化为单期方案虽剂量降低超过60%,但诊断准确度下降,敏感度为83.33%。
研究设计与样本
- 参考检查为完整三期方案:真实平扫(TNC)加动脉期和60秒延迟期多能重建。
- 读片会话II和III由两名独立盲法读者评估简化方案VMI:双期(VNC+动脉期+延迟期)和单期(VNC+延迟期)。
诊断标准与关键数值
- 双期VMI方案检出内漏的敏感度为100%。
- 单期VMI方案检出内漏的敏感度为83.33%。
- 双期方案平均有效剂量降低23.28%,单期方案降低61.37%。
临床应用与阅片要点
- EVAR术后随访可采用双期VMI方案(VNC+动脉期+延迟期)替代传统三期CT,敏感度高且剂量更低。
- 单期方案虽剂量更低,但敏感度不足,可能漏诊内漏,需谨慎选择。
局限性
- 未报告特异度、AUC等完整诊断效能指标。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The objective of this prospective study was to evaluate the virtual monoenergetic images (VMI) and virtual noncontrast (VNC) phase in the detection of endoleaks after endovascular abdominal aortic repair (EVAR). The potential dose reduction of abbreviated examination protocols was calculated.
MATERIALS AND METHODS: Ninety-seven patients after the EVAR procedure were enrolled in this study. An initial single-source noncontrast acquisition was followed by two dual-energy acquisitions (arterial and 60 s delayed). Fast-kVp switching scanner was used. VNC images were reconstructed from the delayed phase. First examination session (reference) included a full triphasic study protocol consisting of true noncontrast (TNC) images and two postcontrast phases, the latter ones presented as classical polyenergetic reconstructions. Reading sessions II and III were performed by two independent and blinded readers evaluating VMIs in abbreviated protocols-biphasic (VNC + arterial, delayed phase), monophasic (VNC + delayed phase). The diagnostic accuracy of sessions II and III was calculated.
RESULTS: The calculated sensitivity of the biphasic protocol with the use of VMIs in endoleak detection was 100%, with a statistically significant increase in the number of endoleaks detected in comparison with the reference study. The monophasic protocol showed 83.33% sensitivity. The use of abbreviated examination protocols led to a decrease in the mean effective dose (ED) of 23.28% (biphasic protocol) and 61.37% (monophasic protocol).
CONCLUSION: The use of VMIs increases the number of endoleaks diagnosed with a possible radiation reduction by up to ¼ (biphasic protocol). Further reduction to a monophasic protocol leads to over 60% dose reduction but with a decrease in diagnostic accuracy.