双层探测器锥形束CT血管造影用于卒中评估:首次人体试验结果
排除运动或对比剂注射问题后,读者认为该技术在评估取栓候选动脉时非劣效于CTA,但伪影更多,且扫描时间长、无法团注追踪。
要点
- 双层探测器锥形束CTA在排除扫描问题后对取栓相关动脉的显示非劣效于CTA。
- 该技术伪影较CTA更常见,除M1段外多数节段清晰度非劣效。
- 原型设备扫描时间长且无法对比剂团注追踪,可能限制其急性卒中应用。
关键影像表现/征象
- 双层探测器锥形束CTA上伪影较CTA更常见。
- 多数评估认为除M1段外各节段清晰度非劣效于CTA。
诊断标准与关键数值
- 非劣效性判定采用精确二项检验,单侧下限性能边界预设为80%(98.75% CI)。
- 排除运动或对比剂注射问题后,读者个体评估的CI边界分别为93%、84%、80%,均认为双层探测器锥形束CTA非劣效于CTA。
临床应用与阅片要点
- 在取栓候选患者中,可考虑使用双层探测器锥形束CTA评估颅内动脉,但需注意伪影可能影响判读。
- 扫描时间长和无法团注追踪可能影响图像质量,阅片时应关注运动或对比剂注射相关问题。
局限性
- 原型设备扫描时间长且无法进行对比剂团注追踪。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: In patients with stroke, IV cone-beam CTA in the angiography suite could be an alternative to CTA to shorten the door-to-thrombectomy time. However, image quality in cone-beam CTA is typically limited by artifacts. This study evaluated a prototype dual-layer detector cone-beam CT angiography versus CTA in patients with stroke.
MATERIALS AND METHODS: A prospective, single-center trial enrolled consecutive patients with ischemic or hemorrhagic stroke on initial CT. Intracranial arterial segment vessel conspicuity and artifact presence were evaluated on dual-layer cone-beam CTA 70-keV virtual monoenergetic images and CTA. Eleven predetermined vessel segments were matched for every patient. Twelve patients were necessary to show noninferiority to CTA. Noninferiority was determined by the exact binomial test; the 1-sided lower performance boundary was prospectively set to 80% (98.75% CI).
RESULTS: Twenty-one patients had matched image sets (mean age, 72 years). After excluding examinations with movement or contrast media injection issues, all readers individually considered dual-layer cone-beam CT angiography noninferior to CTA (CI boundary, 93%, 84%, 80%, respectively) when evaluating arteries relevant in candidates for intracranial thrombectomy. Artifacts were more prevalent compared with CTA. The majority assessment rated each individual segment except M1 as having noninferior conspicuity compared with CTA.
CONCLUSIONS: In a single-center stroke setting, dual-layer detector cone-beam CTA virtual monoenergetic images are noninferior to CTA under certain conditions. Notably, the prototype is hampered by a long scan time and is not capable of contrast media bolus tracking. After excluding examinations with such scan issues, readers considered dual-layer detector cone-beam CTA noninferior to CTA, despite more artifacts.
引用
- Ståhl, Almqvist, Kolloch, et al. Dual-Layer Detector Cone-Beam CT Angiography for Stroke Assessment: First-in-Human Results (the Next Generation X-ray Imaging System Trial). AJNR Am J Neuroradiol. 2023. PMID: 37055159. doi:10.3174/ajnr.A7835
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。