利用焦点沸腾气泡原位测量聚焦超声消融手术的声衰减
研究提出一种基于焦点沸腾气泡作为强反射体的原位声衰减测量方法,并在离体猪里脊和牛心脏中评估。在0.97 MHz、3 cm厚度下测得衰减系数分别为0.159±0.002和0.250±0.005 Np/cm,与文献范围一致。该方法有望临床转化以提高聚焦超声消融的安全性和有效性。
要点
- 焦点沸腾气泡可作为强反射体用于原位声衰减测量。
- 离体猪里脊和牛心脏的衰减系数与文献值一致。
- 该方法操作简便,有望临床转化。
关键影像表现/征象
- 焦点处产生数毫米大小的沸腾气泡作为强反射体。
- 回声振幅对传播路径条件敏感。
诊断标准与关键数值
- 0.97 MHz、3 cm厚度下,猪里脊声衰减系数为0.159±0.002 Np/cm。
- 0.97 MHz、3 cm厚度下,牛心脏声衰减系数为0.250±0.005 Np/cm。
- 硅胶垫反向声衰减系数为0.807±0.002 Np/cm,与插入替代法0.766±0.003 Np/cm相当。
临床应用与阅片要点
- 可在聚焦超声消融术前评估组织声衰减,辅助治疗计划。
- 注意回声振幅受传播路径条件影响。
局限性
- 仅离体实验验证,未进行在体或临床研究。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Acoustic attenuation in the propagation path of focused ultrasound ablation surgery determines the energy loss toward the focal region and is critical to the consequent treatment outcomes. In situ non-invasive, reliable, and accurate measurement is challenging for multi-layered heterogeneous tissues within the focusing angle.
METHODS: A novel measurement approach is proposed and its performance is evaluated using ex vivo porcine tenderloin and bovine heart. A big boiling bubble (i.e., larger than a few millimeters in size) was produced at the focus as a strong reflector inside the tissue, and the echo amplitudes were used to determine the acoustic attenuation. Two models, acoustic ray and energy loss, were developed to derive the equivalent acoustic attenuation coefficient for a focused beam.
RESULTS: The measured acoustic attenuation coefficients of ex vivo porcine tenderloin and bovine heart at 0.97 MHz and a thickness of 3 cm are 0.159 ± 0.002 and 0.250 ± 0.005 Np/cm, respectively, which are all within the scope of measured values in the literature. In addition, the echo amplitude is sensitive to the conditions of the propagation path, and the inverse acoustic attenuation coefficient of the silicone gel pad placed in front of the tissue sample was 0.807 ± 0.002 Np/cm, which is comparable to the measurement using the insertion substitution method, 0.766 ± 0.003 Np/cm.
CONCLUSION: Our proposed approach could determine the tissue acoustic attenuation for focused ultrasound ablation surgery reliably and accurately in situ. The easy operating protocol may allow clinical translation and adoption for improved safety and efficacy.
引用
- Zhang, Zhou, Wang, et al. In Situ Measurement of Acoustic Attenuation for Focused Ultrasound Ablation Surgery Using a Boiling Bubble at the Focus. Ultrasound Med Biol. 2023. PMID: 37005115. doi:10.1016/j.ultrasmedbio.2023.02.014
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。