文献速递神经技术与安全CT

深度学习重建可提升低剂量头颅CT图像质量

Deep learning-based reconstruction can improve the image quality of low radiation dose head CT.
PubMed 文献 · Eur Radiol · 2023-03-28

本研究回顾性纳入114例接受标准剂量或低剂量头颅平扫CT的患者,比较混合迭代重建、基于模型迭代重建与深度学习重建的图像质量。低剂量方案有效剂量降低25%,其中深度学习重建在降低噪声、提高灰白质对比和病灶显示方面优于混合迭代重建,且重建时间较短。基于模型迭代重建虽降低噪声但噪声纹理、锐利度和主观接受度较差且重建时间长。

要点

  • 低剂量头颅CT采用深度学习重建可在降低25%剂量下获得优于标准剂量混合迭代重建的图像质量。
  • 深度学习重建的病灶显示评分(2.9)高于混合迭代重建(1.2)和基于模型迭代重建(1.8)。
  • 基于模型迭代重建噪声纹理和锐利度较差且重建时间长达319秒,临床可行性受限。

局限性

  • 本研究为回顾性单中心设计,样本量有限。
  • 未报告病灶显示评估的具体病种和参考标准。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To evaluate the image quality of deep learning-based reconstruction (DLR), model-based (MBIR), and hybrid iterative reconstruction (HIR) algorithms for lower-dose (LD) unenhanced head CT and compare it with those of standard-dose (STD) HIR images.

METHODS: This retrospective study included 114 patients who underwent unenhanced head CT using the STD (n = 57) or LD (n = 57) protocol on a 320-row CT. STD images were reconstructed with HIR; LD images were reconstructed with HIR (LD-HIR), MBIR (LD-MBIR), and DLR (LD-DLR). The image noise, gray and white matter (GM-WM) contrast, and contrast-to-noise ratio (CNR) at the basal ganglia and posterior fossa levels were quantified. The noise magnitude, noise texture, GM-WM contrast, image sharpness, streak artifact, and subjective acceptability were independently scored by three radiologists (1 = worst, 5 = best). The lesion conspicuity of LD-HIR, LD-MBIR, and LD-DLR was ranked through side-by-side assessments (1 = worst, 3 = best). Reconstruction times of three algorithms were measured.

RESULTS: The effective dose of LD was 25% lower than that of STD. Lower image noise, higher GM-WM contrast, and higher CNR were observed in LD-DLR and LD-MBIR than those in STD (all, p ≤ 0.035). Compared with STD, the noise texture, image sharpness, and subjective acceptability were inferior for LD-MBIR and superior for LD-DLR (all, p < 0.001). The lesion conspicuity of LD-DLR (2.9 ± 0.2) was higher than that of HIR (1.2 ± 0.3) and MBIR (1.8 ± 0.4) (all, p < 0.001). Reconstruction times of HIR, MBIR, and DLR were 11 ± 1, 319 ± 17, and 24 ± 1 s, respectively.

CONCLUSION: DLR can enhance the image quality of head CT while preserving low radiation dose level and short reconstruction time.

KEY POINTS: • For unenhanced head CT, DLR reduced the image noise and improved the GM-WM contrast and lesion delineation without sacrificing the natural noise texture and image sharpness relative to HIR. • The subjective and objective image quality of DLR was better than that of HIR even at 25% reduced dose without considerably increasing the image reconstruction times (24 s vs. 11 s). • Despite the strong noise reduction and improved GM-WM contrast performance, MBIR degraded the noise texture, sharpness, and subjective acceptance with prolonged reconstruction times relative to HIR, potentially hampering its feasibility.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Nagayama, Iwashita, Maruyama, et al. Deep learning-based reconstruction can improve the image quality of low radiation dose head CT. Eur Radiol. 2023. PMID: 36973431. doi:10.1007/s00330-023-09559-3

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。