高分辨血管壁MR联合TOF-MRA随访颅内椎基底动脉夹层动脉瘤重建性血管内治疗后的诊断效能
本研究以DSA为参考标准,评估3T高分辨血管壁MR联合TOF-MRA在重建性血管内治疗后随访颅内椎基底动脉夹层动脉瘤的诊断效能。
要点
- 联合TOF-MRA与高分辨血管壁MR可提高动脉瘤残留检出的敏感度至93.3%,特异度达100.0%。
- 高分辨血管壁MR对支架血管显示评分(4.88)显著优于TOF-MRA(2.53),P<.001。
- 高分辨血管壁MR评估载瘤动脉通畅性与DSA完全一致,TOF-MRA特异度仅8.0%。
关键影像表现/征象
- 高分辨血管壁MR对支架血管的显示评分均值为4.88(SD,0.32),显著高于TOF-MRA的2.53(SD,1.25),P<.001。
- 高分辨血管壁MR评估载瘤动脉通畅性(正常或异常)与DSA完全一致。
诊断标准与关键数值
- TOF-MRA诊断动脉瘤残留的敏感度80.0%、特异度100.0%、阳性预测值100.0%、阴性预测值82.4%。
- 高分辨血管壁MR诊断动脉瘤残留的敏感度53.3%、特异度100.0%、阳性预测值100.0%、阴性预测值66.7%。
- 高分辨血管壁MR联合TOF-MRA诊断动脉瘤残留的敏感度93.3%、特异度100.0%、阳性预测值100.0%、阴性预测值93.3%。
- TOF-MRA评估载瘤动脉通畅性的敏感度100.0%、特异度8.0%、阳性预测值14.8%、阴性预测值100.0%。
临床应用与阅片要点
- 随访中联合TOF-MRA与高分辨血管壁MR可提高动脉瘤残留检出敏感度并保持高特异度。
- 高分辨血管壁MR能清晰显示支架血管,有助于评估载瘤动脉通畅性,弥补TOF-MRA在支架内评估的不足。
局限性
- 为单中心研究,结论外推需谨慎。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Few studies have reported the utility of high-resolution vessel wall MR imaging in the follow-up of endovascularly treated vertebrobasilar dissecting aneurysms. This study aimed to evaluate the diagnostic performance of high-resolution vessel wall MR imaging combined with TOF-MRA in the follow-up of intracranial vertebrobasilar dissecting aneurysms after reconstructive endovascular treatment.
MATERIALS AND METHODS: Patients with intracranial vertebrobasilar dissecting aneurysms with reconstructive endovascular treatment and followed up with TOF-MRA, high-resolution vessel wall MR imaging, and DSA were included. With DSA as the criterion standard, the diagnostic performance of TOF-MRA, high-resolution vessel wall MR imaging, and high-resolution vessel wall MR imaging combined with TOF-MRA in the evaluation of aneurysm occlusion status and parent artery patency was assessed. Visualization of the stented artery on TOF-MRA and high-resolution vessel wall MR imaging was rated on a 5-point scale.
RESULTS: Twenty-seven patients with 29 aneurysms were included. The sensitivity, specificity, positive predictive value, and negative predictive value of TOF-MRA, high-resolution vessel wall MR imaging, and high-resolution vessel wall MR imaging combined with TOF-MRA for diagnosing aneurysm remnants were 80.0%, 100.0%, 100.0%, and 82.4%; 53.3%, 100.0%, 100.0%, and 66.7%; and 93.3%, 100.0%, 100.0%, and 93.3%, respectively. For the visualization of the stented artery, the mean score of high-resolution vessel wall MR imaging was significantly higher than that of TOF-MRA (4.88 [SD, 0.32] versus 2.53 [SD, 1.25], P < .001). In the evaluation of parent artery patency (normal or pathologic), whereas TOF-MRA had a sensitivity, specificity, positive predictive value, and negative predictive value of 100.0%, 8.0%, 14.8%, and 100.0%, respectively, high-resolution vessel wall MR imaging was completely consistent with the DSA.
CONCLUSIONS: High-resolution vessel wall MR imaging combined with TOF-MRA at 3T showed good diagnostic performance in the follow-up of intracranial vertebrobasilar dissecting aneurysms after reconstructive endovascular treatment.
引用
- Tan, Zhou, Xu, et al. Diagnostic Performance of High-Resolution Vessel Wall MR Imaging Combined with TOF-MRA in the Follow-up of Intracranial Vertebrobasilar Dissecting Aneurysms after Reconstructive Endovascular Treatment. AJNR Am J Neuroradiol. 2023. PMID: 36958804. doi:10.3174/ajnr.A7838
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。