三维超声测量腘绳肌与肌腱体积的效度与信度
本研究纳入13名参与者,在两天内进行两次自由臂三维超声采集,并与MRI对照,评估腘绳肌各头及股薄肌的肌肉体积和半腱肌、股薄肌肌腱体积的效度与信度。结果显示三维超声与MRI的偏差和95%置信区间在肌肉体积为-1.9 mL(-0.8%)至1.2 mL(1.0%),肌腱体积为0.01 mL(0.2%)至-0.03 mL(-2.6%);肌肉体积的ICC为0.98(股薄肌)至1.00,CV为1.1%(半膜肌)至3.4%(股二头肌短头);肌腱体积的ICC为0.99,CV为3.2%(半腱肌腱)至3.4%(股薄肌腱)。结论认为三维超声可提供有效的、可靠的隔天测量,未来可作为强化干预的结果指标并可能用于临床环境。
要点
- 三维超声测量腘绳肌和股薄肌肌肉体积与MRI偏差小,ICC高(0.98-1.00)。
- 肌腱体积测量同样可靠,ICC为0.99,CV为3.2%-3.4%。
- 三维超声可用于肌肉肌腱体积的纵向监测,但需注意设备与操作标准化。
局限性
- 样本量较小(13名参与者),可能影响结果的普遍性。
- 仅评估了腘绳肌和股薄肌,未涉及其他肌肉。
- 研究设计未明确提及盲法或随机化。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The validity and reliability of 3-D ultrasound (US) in estimation of muscle and tendon volume was assessed in a very limited number of muscles that can be easily immersed. The objective of the present study was to assess the validity and reliability of muscle volume measurements for all hamstring muscle heads and gracilis (GR), as well as tendon volume for the semitendinosus (ST) and GR using freehand 3-D US.
METHODS: Three-dimensional US acquisitions were performed for 13 participants in two distinct sessions on separate days, in addition to one session dedicated to magnetic resonance imaging (MRI). Volumes of ST, semimembranosus (SM), biceps femoris short (BFsh) and long (BFlh) heads, and GR muscles and from the tendon from semitendinosus (STtd) and gracilis (GRtd) were collected.
RESULTS: The bias and the 95% confidence intervals of 3-D US compared with MRI ranged from -1.9 mL (-0.8%) to 1.2 mL (1.0%) for muscle volume and from 0.01 mL (0.2%) to -0.03 mL (-2.6%) for tendon volume. For muscle volume assessed using 3-D US, intraclass correlation coefficients (ICCs) ranged from 0.98 (GR) to 1.00, and coefficients of variation (CV) from 1.1% (SM) to 3.4% (BFsh). For tendon volume, ICCs were 0.99, and CVs between 3.2% (STtd) and 3.4% (GRtd).
CONCLUSION: Three-dimensional US can provide a valid and reliable inter-day measurement of hamstrings and GR for both muscle and tendon volumes. In the future, this technique could be used as an outcome for strengthening interventions and potentially in clinical environments.