超声与纤维喉镜确认喉罩置入的比较
前瞻性观察研究纳入100例ASA I-II级全麻手术患者,临床确认喉罩置入后分别用超声和纤维喉镜评估可接受性。超声与纤维喉镜在85%患者中一致(κ=0.690),超声诊断准确率85%,纤维喉镜评估不可接受者均更换气管导管。
要点
- 超声可用于确认喉罩位置,与纤维喉镜一致率85%。
- 纤维喉镜评估不可接受者均需更换气管导管。
- 超声提供安全、无创、实时动态评估。
局限性
- 单中心前瞻性观察研究,样本量100例。
- 未报告超声评估的具体影像标准。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: 1) To compare ultrasound (US) examination and fiberoptic laryngoscopy (FOL) for confirmation of laryngeal mask airway (LMA) placement. 2) To evaluate the necessity for reinsertion of LMA based on FOL.
METHODS: This prospective observational study included 100 adult patients of American Society of Anesthesiologists (ASA) Grade I and II, undergoing elective surgery under General Anesthesia requiring Proseal LMA™ placement as an airway device. LMA placement was first confirmed by clinical tests. Clinically acceptable patients were further assessed by US and categorized as acceptable (US-A) or unacceptable (US-U) and again by FOL as (FOL-A and FOL-U). Categorical variables presented in number, percentage (%), and continuous variables presented as mean ± SD and median. Inter-rater kappa agreement was used to find out the strength of agreement of acceptability between FOL and US.
RESULTS: The LMA placement was clinically acceptable in 82% of patients on first attempt. FOL had 63% (FOL-A) acceptable LMA placement as compared with US examination which had 56% (US-A). In 85% of patients, US and FOL findings were in good agreement with each other for LMA placement (κ = 0.690 and P < .05). In all patients of FOL of unacceptable (FOL-U) category (37%), LMA was replaced with endotracheal tube.
CONCLUSION: US provides a safe, non-invasive, and real-time dynamic assessment with 85% diagnostic accuracy for confirmation of LMA placement as compared with FOL.