光子计数CT与临床MRI肝脏脂肪定量头对头比较的首次经验

Liver fat quantification in photon counting CT in head to head comparison with clinical MRI - First experience.
PubMed 文献 · Eur J Radiol · 2023-02-09

本研究通过体模和12例肥胖脂肪肝患者,比较光子计数CT(PCCT)与1.5T MRI(IDEAL-IQ序列)的肝脏脂肪定量。体模中PCCT测得脂肪和碘浓度与已知值无显著差异;患者中MRI与PCCT平均脂肪信号分数分别为13.1±9.9%和12.0±9.0%,平均差异1.1±1.9%(P=0.138)。结论认为PCCT在肥胖患者肝脏脂肪定量中显示出有前景的准确性,未来可能改善CT的机会性筛查。

要点

  • PCCT与MRI肝脏脂肪分数平均差异仅1.1±1.9%,两者定量一致性良好。
  • 体模验证PCCT可同时准确量化脂肪和碘,不受碘对比剂影响。
  • PCCT脂肪定量有望用于腹部CT的机会性脂肪肝筛查。

局限性

  • 样本量仅12例,为首次经验,结论需更大样本验证。
  • 仅纳入肥胖患者,未覆盖非肥胖人群。
  • 体模与临床测量均来自单一中心、单一设备型号。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To compare liver fat quantification between MRI and photon-counting CT (PCCT).

METHOD: A cylindrical phantom with inserts containing six concentrations of oil (0, 10, 20, 30, 50 and 100%) and oil-iodine mixtures (0, 10, 20, 30 and 50% fat +3 mg/mL iodine) was imaged with a PCCT (NAEOTOM Alpha) and a 1.5 T MRI system (MR 450w, IDEAL-IQ sequence), using clinical parameters. An IRB-approved prospective clinical evaluation included 12 obese adult patients with known fatty liver disease (seven women, mean age: 61.5 ± 13 years, mean BMI: 30.3 ± 4.7 kg/m2). Patients underwent a same-day clinical MRI and PCCT of the abdomen. Liver fat fractions were calculated for four segments (I, II, IVa and VII) using in- and opposed-phase on MRI ((Meanin - Meanopp)/2*Meanin) and iodine-fat, tissue decomposition analysis in PCCT (Syngo.Via VB60A). CT and MRI Fat fractions were compared using two-sample t-tests with equal variance. Statistical analysis was performed using RStudio (Version1.4.1717).

RESULTS: Phantom results showed no significant differences between the known fat fractions (P = 0.32) or iodine (P = 0.6) in comparison to PCCT-measured concentrations, and no statistically significant difference between known and MRI-measured fat fractions (P = 0.363). In patients, the mean fat signal fraction measured on MRI and PCCT was 13.1 ± 9.9% and 12.0 ± 9.0%, respectively, with an average difference of 1.1 ± 1.9% between the modalities (P = 0.138).

CONCLUSION: First experience shows promising accuracy of liver fat fraction quantification for PCCT in obese patients. This method may improve opportunistic screening for CT in the future.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Schwartz, Ashton, Wildman-Tobriner, et al. Liver fat quantification in photon counting CT in head to head comparison with clinical MRI - First experience. Eur J Radiol. 2023. PMID: 36842273. doi:10.1016/j.ejrad.2023.110734

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。