低频透射超声全身成像:三维超声断层扫描的定量精度与MRI对照研究
本研究在4只10-12日龄猪仔中,以3T MRI和大体解剖为参考标准,评估基于约1 MHz低频透射超声和3.6 MHz反射数据的3D超声断层扫描(volography)全身成像的声速准确性和图像质量。结果显示组织声速值与文献值平均偏差1.07%(0.5%),图像对比度和分辨率优于3T MRI,可显示肾髓质、皮质、纤维被膜及回肠肠腔细节。
要点
- volography可在存在骨和空气的情况下获得高分辨率、定量准确的全身图像。
- 组织声速测量值与文献值平均偏差仅1.07%(0.5%),支持定量成像潜力。
- 图像对比度和分辨率优于3T MRI,可分辨肾脏分层结构和肠腔细节。
关键影像表现/征象
- volography图像在轴位、冠状位和矢状位上与MRI详细对应。
- 图像显示肾脏髓质、皮质和纤维被膜分层结构。
- 图像显示小肠回肠肠腔细节。
- volography图像对比度和分辨率优于3T MRI。
诊断标准与关键数值
- 组织声速值与文献值平均偏差1.07%(标准差0.5%)。
- 反射数据采用三维折射校正射线追踪,实现360°复合和亚毫米分辨率。
临床应用与阅片要点
- volography可作为全身介观尺度成像的安全、低成本、高分辨率模态。
- 阅片时注意其定量声速信息可与MRI形态学互补。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: To indicate that 3D low-frequency ultrasound tomography with 3D data acquisition (volography) is a safe, low-cost, high-resolution, whole-body meso-scale medical imaging modality that gives high-resolution quantitatively accurate clinically relevant images.
MATERIALS AND METHODS: We compare the speed of sound accuracy in various organs in situ. We validate our 3D ultrasound tomography images using MRI and gross section anatomy as ground truth in 10-day old piglets. Data acquisition is accomplished with the QT Scanner at ∼1 MHz center frequency, and array transceivers for reflection data @3.6 MHz. Images are generated with unique model-based 3D ultrasound tomography algorithms. In reflection, we use 3D refraction-corrected ray tracing to allow 360° compounding with sub-mm resolution. Four 10-12 day old pigs were anesthetized and whole-body images were acquired via low-frequency transmitted ultrasound and 3T MRI.
RESULTS: Tissue values were within an average of 1.07% (0.5%) of the literature values. We also show the detailed correlation of our images with MRI images in axial, coronal, and sagittal views. Volography images of a piglet show high resolution and quantitative accuracy, showing more contrast &resolution than 3T MRI, including the kidney showing medulla, cortex and fibrous cover, and small intestines with ileal lumen detail visible.
CONCLUSION: We establish that 3D ultrasound tomography (volography), yields high-resolution quantitatively accurate images whole-body images in presence of bone and air which are potentially clinically useful but have not appeared in the literature.