磁共振测温靶向用于磁共振引导组织碎化术治疗
本研究在离体牛脑样本中评估超声诱导低温加热联合磁共振测温为经颅磁共振引导组织碎化术提供治疗前靶向定位的可行性与准确性。七例样本先加热约1.6°C并用磁共振测温定位靶点,随后生成组织碎化病灶,横向和纵向定位误差分别为0.59/0.31 mm和1.31/0.93 mm。结果表明磁共振测温可为经颅磁共振引导组织碎化术提供可靠的预处理靶向。
要点
- 磁共振测温可在组织碎化前定位靶点,横向误差约0.59 mm。
- 纵向定位误差约1.31 mm,提示纵向精度需关注。
- 该方法在离体牛脑验证,尚未在体应用。
核心结论
- 磁共振测温可为经颅磁共振引导组织碎化术提供可靠的治疗前靶向定位。
研究设计与样本
- 离体牛脑样本研究,共7例样本。
- 使用15阵元、750 kHz磁共振兼容超声换能器,可输出低温加热与组织碎化脉冲。
- 参考标准为组织碎化后磁共振图像上病灶的质心。
关键影像表现/征象
- 磁共振测温通过焦点处约1.6°C的温升定位靶点。
- 组织碎化病灶在治疗后磁共振图像上可见。
诊断标准与关键数值
- 磁共振测温峰值加热点与治疗后组织碎化病灶质心的差值:横向0.59/0.31 mm,纵向1.31/0.93 mm(均值/标准差)。
临床应用与阅片要点
- 治疗前可利用约1.6°C的低温加热联合磁共振测温确认靶点位置。
- 关注纵向定位误差大于横向,报告时应描述靶向精度。
局限性
- 研究为离体牛脑样本,非在体脑组织。
- 样本量较小(7例)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The potential of transcranial magnetic resonance (MR)-guided histotripsy for brain applications has been described in prior in vivo studies in the swine brain through an excised human skull. The safety and accuracy of transcranial MR-guided histotripsy (tcMRgHt) rely on pre-treatment targeting guidance. In the work described here, we investigated the feasibility and accuracy of using ultrasound-induced low-temperature heating and MR thermometry for histotripsy pre-treatment targeting in ex vivo bovine brain.
METHODS: A 15-element, 750-kHz MRI-compatible ultrasound transducer with modified drivers that can deliver both low-temperature heating and histotripsy acoustic pulses was used to treat seven bovine brain samples. The samples were first heated to an approximately 1.6°C temperature increase at the focus, and MR thermometry was used to localize the target. Once the targeting was confirmed, a histotripsy lesion was generated at the focus and visualized on post-histotripsy MR images.
DISCUSSION: The accuracy of MR thermometry targeting was evaluated with the mean/standard deviation of the difference between the locus of peak heating identified by MR thermometry and the center of mass of the post-treatment histotripsy lesion, which was 0.59/0.31 mm and 1.31/0.93 mm in the transverse and longitudinal directions, respectively.
CONCLUSION: This study determined that MR thermometry could provide reliable pre-treatment targeting for transcranial MR-guided histotripsy treatment.