文献速递骨肌MR

深度学习重建加速手与腕部TSE成像:图像质量、解剖显示与常见病变检测对比

Deep Learning MRI Reconstruction for Accelerating Turbo Spin Echo Hand and Wrist Imaging: A Comparison of Image Quality, Visualization of Anatomy, and Detection of Common Pathologies with Standard Imaging.
PubMed 文献 · Acad Radiol · 2023-02-14

该前瞻性研究纳入21例患者,对比标准全采样GRAPPA重建TSE(TSES)与前瞻性欠采样深度学习重建TSE(TSEDL)在手与腕部MRI中的表现。结果显示TSEDL在整体图像质量、伪影、解剖结构显示及诊断信心上与TSES相当,且噪声更低、边缘更锐利,采集时间缩短超过60%。

要点

  • TSEDL在保持诊断质量的同时将手/腕TSE序列采集时间缩短60%以上
  • TSEDL较TSES噪声更低、边缘更锐利,整体图像质量与诊断信心相当
  • 对三角纤维软骨复合体、肌腱、韧带、正中神经及软骨的显示与TSES无显著差异

关键影像表现/征象

  • TSEDL整体图像质量、伪影、解剖结构显示及诊断信心与TSES相当(p > 0.05)。
  • TSEDL图像噪声评分(4.90,中位数5,IQR 5-5)优于TSES(4.52,中位数5,IQR 4-5),p < 0.05。
  • TSEDL边缘锐利度评分(4.10,中位数4,IQR 3.5-5)优于TSES(3.57,中位数4,IQR 3-4),p < 0.05。
  • 评估的解剖结构包括三角纤维软骨复合体、尺侧腕伸肌腱、外在及内在韧带、正中神经和软骨。

诊断标准与关键数值

  • 采用五点Likert量表定性评估图像质量、噪声、边缘锐利度、伪影、诊断信心及解剖结构显示。
  • 常见病变检测的读者间及读者内一致性为实质性至几乎完全一致(κ = 0.632-1.000)。

临床应用与阅片要点

  • 在日常手/腕MRI中可采用深度学习重建TSE序列以大幅缩短扫描时间,同时保持诊断质量。
  • 深度学习重建可能降低运动伪影风险并提高患者舒适度和检查通量。
  • 阅片时注意TSEDL图像噪声更低、边缘更锐利,但整体诊断信心与标准序列相当。

局限性

  • 样本量较小(21例),可能限制统计效能和泛化性。
  • 单中心前瞻性研究,需多中心验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: Magnetic resonance imaging (MRI) of the hand and wrist is a routine MRI examination and takes about 15-20 minutes, which can lead to problems resulting from the relatively long scan time, such as decreased image quality due to motion artifacts and lower patient throughput. The objective of this study was to evaluate a deep learning (DL) reconstruction for turbo spin echo (TSE) sequences of the hand and wrist regarding image quality, visualization of anatomy, and diagnostic performance concerning common pathologies.

MATERIALS AND METHODS: Twenty-one patients (mean age: 43 ± 19 [19-85] years, 10 men, 11 female) were prospectively enrolled in this study between October 2020 and June 2021. Each participant underwent two MRI protocols: first, standard fully sampled TSE sequences reconstructed with a standard GRAPPA reconstruction (TSES) and second, prospectively undersampled TSE sequences using a conventional parallel imaging undersampling pattern reconstructed with a DL reconstruction (TSEDL). Both protocols were acquired consecutively in one examination. Two experienced MSK-imaging radiologists qualitatively evaluated the images concerning image quality, noise, edge sharpness, artifacts, and diagnostic confidence, as well as the delineation of anatomical structures (triangular fibrocartilage complex, tendon of the extensor carpi ulnaris muscle, extrinsic and intrinsic ligaments, median nerve, cartilage) using a five-point Likert scale and assessed common pathologies. Wilcoxon signed-rank test and kappa statistics were performed to compare the sequences.

RESULTS: Overall image quality, artifacts, delineation of anatomical structures, and diagnostic confidence of TSEDL were rated to be comparable to TSES (p > 0.05). Additionally, TSEDL showed decreased image noise (4.90, median 5, IQR 5-5) compared to TSES (4.52, median 5, IQR 4-5, p < 0.05) and improved edge sharpness (TSEDL: 4.10, median 4, IQR 3.5-5; TSES: 3.57, median 4, IQR 3-4; p < 0.05). Inter- and intrareader agreement was substantial to almost perfect (κ = 0.632-1.000) for the detection of common pathologies. Time of acquisition could be reduced by more than 60% with the protocol using TSEDL.

CONCLUSION: Compared to TSES, TSEDL provided decreased noise and increased edge sharpness, equal image quality, delineation of anatomical structures, detection of pathologies, and diagnostic confidence. Therefore, TSEDL may be clinically relevant for hand and wrist imaging, as it reduces examination time by more than 60%, thus increasing patient comfort and patient throughput.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Herrmann, Gassenmaier, Keller, et al. Deep Learning MRI Reconstruction for Accelerating Turbo Spin Echo Hand and Wrist Imaging: A Comparison of Image Quality, Visualization of Anatomy, and Detection of Common Pathologies with Standard Imaging. Acad Radiol. 2023. PMID: 36797172. doi:10.1016/j.acra.2022.12.042

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。