文献速递神经MR

自动化脑形态测量工具在初发记忆症状患者中的效度评估

Validity Assessment of an Automated Brain Morphometry Tool for Patients with De Novo Memory Symptoms.
PubMed 文献 · AJNR Am J Neuroradiol · 2023-02-16

本研究从OASIS-4数据库选取45例初发记忆症状患者的T1加权图像,比较AI-Rad Companion脑MR软件与in-house FreeSurfer 7.1.1/ILP流程的脑分割性能。结果显示两者绝对体积强相关但一致性中等、一致性差,标准化后相关性增强;以FreeSurfer为参考标准,AI工具检测体积异常的敏感度64.3%-100%、特异度90.6%-100%,两种工具生成的放射学印象与临床诊断兼容率无差异。

要点

  • AI工具检测脑萎缩敏感度64.3%-100%、特异度90.6%-100%,可用于初发记忆症状的萎缩筛查。
  • AI与FreeSurfer绝对体积强相关但一致性差,标准化后相关性提高,标准化测量值差异显著。
  • 两种工具生成的放射学印象与临床诊断兼容率无差异,支持AI工具辅助报告。

核心结论

  • AI-Rad Companion脑MR工具能可靠检测与痴呆鉴别诊断相关的皮层和皮层下区域萎缩。
  • AI工具与FreeSurfer在绝对体积上强相关、一致性中等、一致性差,标准化后相关性增强,标准化测量值差异显著。
  • 以FreeSurfer 7.1.1/ILP为参考标准,AI工具检测体积异常的敏感度为64.3%-100%、特异度为90.6%-100%。
  • 使用两种工具时,放射学印象与临床诊断的兼容率无差异。

研究设计与样本

  • 研究为回顾性方法学比较,样本为45例初发记忆症状患者,图像选自OASIS-4数据库。
  • 使用T1加权图像,分别经AI-Rad Companion脑MR软件和in-house FreeSurfer 7.1.1/Individual Longitudinal Participant流程分析。
  • 参考标准为FreeSurfer 7.1.1/Individual Longitudinal Participant流程。
  • 比较两种工具在绝对、归一化和标准化体积上的相关性、一致性和一致性。
  • 使用各工具生成的最终报告比较异常检出率及放射学印象与临床诊断的兼容性。

诊断标准与关键数值

  • 以FreeSurfer 7.1.1/Individual Longitudinal Participant流程为参考标准,AI-Rad Companion脑MR工具检测体积异常的敏感度为64.3%-100%、特异度为90.6%-100%。
  • AI工具与FreeSurfer在主要皮层脑叶和皮层下结构的绝对体积上呈强相关、中等一致性、差一致性。
  • 将测量值归一化至总颅内体积后,两种工具的相关性强度增加。
  • 两种工具的标准化测量值存在显著差异,可能源于各自校准所用的规范数据集不同。

临床应用与阅片要点

  • AI工具可用于初发记忆症状患者脑萎缩的自动定量筛查,高特异度有助于排除体积异常。
  • 阅片时注意AI工具与FreeSurfer绝对体积一致性差,连续随访或跨工具比较应使用归一化或标准化体积。
  • 标准化测量值因规范数据集不同而存在系统差异,解读时应结合工具特定参考范围,避免直接互换阈值。
  • AI工具生成的报告与临床诊断兼容率与FreeSurfer无差异,可作为放射科报告的有效辅助。

局限性

  • 样本量较小(45例),可能影响统计效能和泛化能力。
  • 研究为回顾性设计,图像选自OASIS-4数据库,存在选择偏倚可能。
  • 未报告多中心外部验证,结论推广需谨慎。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Automated volumetric analysis of structural MR imaging allows quantitative assessment of brain atrophy in neurodegenerative disorders. We compared the brain segmentation performance of the AI-Rad Companion brain MR imaging software against an in-house FreeSurfer 7.1.1/Individual Longitudinal Participant pipeline.

MATERIALS AND METHODS: T1-weighted images of 45 participants with de novo memory symptoms were selected from the OASIS-4 database and analyzed through the AI-Rad Companion brain MR imaging tool and the FreeSurfer 7.1.1/Individual Longitudinal Participant pipeline. Correlation, agreement, and consistency between the 2 tools were compared among the absolute, normalized, and standardized volumes. Final reports generated by each tool were used to compare the rates of detection of abnormality and the compatibility of radiologic impressions made using each tool, compared with the clinical diagnoses.

RESULTS: We observed strong correlation, moderate consistency, and poor agreement between absolute volumes of the main cortical lobes and subcortical structures measured by the AI-Rad Companion brain MR imaging tool compared with FreeSurfer. The strength of the correlations increased after normalizing the measurements to the total intracranial volume. Standardized measurements differed significantly between the 2 tools, likely owing to differences in the normative data sets used to calibrate each tool. When considering the FreeSurfer 7.1.1/Individual Longitudinal Participant pipeline as a reference standard, the AI-Rad Companion brain MR imaging tool had a specificity of 90.6%-100% and a sensitivity of 64.3%-100% in detecting volumetric abnormalities. There was no difference between the rate of compatibility of radiologic and clinical impressions when using the 2 tools.

CONCLUSIONS: The AI-Rad Companion brain MR imaging tool reliably detects atrophy in cortical and subcortical regions implicated in the differential diagnosis of dementia.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。

引用

  1. Rahmani, Jindal, Raji, et al. Validity Assessment of an Automated Brain Morphometry Tool for Patients with De Novo Memory Symptoms. AJNR Am J Neuroradiol. 2023. PMID: 36797031. doi:10.3174/ajnr.A7790

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。