3特斯拉乳腺扩散加权MRI:水激发联合频谱脂肪饱和改善病灶可见性与图像质量
在83例患者的高分辨率3T乳腺MRI中,比较水激发联合频谱脂肪饱和(WEXfs)与标准SPAIR脂肪抑制的扩散加权成像。WEXfs-DWI在良恶性病灶的主观图像质量和病灶勾勒清晰度上均优于SPAIR-DWI,病灶大小估计更接近增强T1减影图像。
要点
- WEXfs-DWI在b=800和1600 s/mm²下对良恶性病灶的图像质量和勾勒清晰度均优于SPAIR-DWI。
- WEXfs-DWI在b=1600 s/mm²下信噪比、对比噪声比和信号强度比均更优。
- WEXfs-DWI测量的病灶长径与增强T1减影图像的差异更小。
核心结论
- WEXfs-DWI的主观和客观图像质量均优于标准SPAIR-DWI,对良恶性病灶大小的估计更准确。
研究设计与样本
- 前瞻性采用两种回波平面DWI序列的多参数乳腺MRI研究,纳入83例患者(50.1±12.6岁),设备为3T MRI,参考标准为增强T1减影图像。
关键影像表现/征象
- 三位放射科医师评估b=800和1600 s/mm²及ADC图上88个局灶性病灶(28个恶性,60个良性)的DWI整体质量和病灶勾勒清晰度。
- 无论b值如何,WEXfs-DWI对良恶性病灶的图像质量(所有p<0.001)和病灶勾勒清晰度(所有p≤0.003)均被认为优于SPAIR-DWI。
- 不论病灶性质,WEXfs-DWI在b=1600 s/mm²下提供更优的信噪比、对比噪声比和信号强度比(所有p≤0.05)。
- WEXfs-DWI与增强T1减影图像之间的病灶大小差异小于SPAIR脂肪抑制(所有p≤0.007)。
诊断标准与关键数值
- 恶性病灶的平均ADC值在WEXfs-DWI上更低(p<0.001)。
- 良性病灶在两种技术之间的ADC值无显著差异(p=0.947)。
临床应用与阅片要点
- WEXfs-DWI在b=800和1600 s/mm²下可改善良恶性乳腺病灶的可见性和勾勒清晰度。
- WEXfs-DWI有助于更准确地估计病灶大小,与增强T1减影图像的一致性更好。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: In breast MRI with diffusion-weighted imaging (DWI), fat suppression is essential for eliminating the dominant lipid signal. This investigation evaluates a combined water-excitation-spectral-fatsat method (WEXfs) versus standard spectral attenuated inversion recovery (SPAIR) in high-resolution 3-Tesla breast MRI.
MATERIALS AND METHODS: Multiparametric breast MRI with 2 echo-planar DWI sequences was performed in 83 patients (50.1 ± 12.6 years) employing either WEXfs or SPAIR for fat signal suppression. Three radiologists assessed overall DWI quality and delineability of 88 focal lesions (28 malignant, 60 benign) on images with b values of 800 and 1600 s/mm2, as well as apparent diffusion coefficient (ADC) maps. For each fat suppression method and b value, the longest lesion diameter was determined in addition to measuring the signal intensity in DWI and ADC value in standardized regions of interest.
RESULTS: Regardless of b values, image quality (all p < 0.001) and lesion delineability (all p ≤ 0.003) with WEXfs-DWI were deemed superior compared to SPAIR-DWI in benign and malignant lesions. Irrespective of lesion characterization, WEXfs-DWI provided superior signal-to-noise, contrast-to-noise and signal-intensity ratios with 1600 s/mm2 (all p ≤ 0.05). The lesion size difference between contrast-enhanced T1 subtraction images and DWI was smaller for WEXfs compared to SPAIR fat suppression (all p ≤ 0.007). The mean ADC value in malignant lesions was lower for WEXfs-DWI (p < 0.001), while no significant ADC difference was ascertained between both techniques in benign lesions (p = 0.947).
CONCLUSION: WEXfs-DWI provides better subjective and objective image quality than standard SPAIR-DWI, resulting in a more accurate estimation of benign and malignant lesion size.
引用
- Sauer, Christner, Schlaiß, et al. Diffusion-weighted Breast MRI at 3 Tesla: Improved Lesion Visibility and Image Quality with a Combination of Water-excitation and Spectral Fat Saturation. Acad Radiol. 2023. PMID: 36764882. doi:10.1016/j.acra.2023.01.014
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。