文献速递技术与安全介入US

利用空化云成核与塌缩声发射进行软组织像差校正以提升组织碎化术效果

Soft Tissue Aberration Correction for Histotripsy Using Acoustic Emissions From Cavitation Cloud Nucleation and Collapse.
PubMed 文献 · Ultrasound Med Biol · 2023-02-08

本研究在离体猪肝经腹壁模型中,比较利用空化云成核与塌缩产生的声发射(ACE)进行软组织像差校正(AC)的效果。结果显示,ACE校正平均可恢复水听器校正所得压力幅度的96%±3%,而无校正时仅为71%±5%;使用≥10个脉冲的中位数时,成核与塌缩校正均能获得>96%的压力幅度,单脉冲校正时塌缩信号更稳健(95%-99%)。

要点

  • ACE像差校正可恢复软组织像差所致的大部分压力幅度损失(平均96% vs 无校正71%)。
  • 塌缩信号在单脉冲校正中表现稳健(95%-99%水听器校正压力),而成核信号变异较大(49%-99%)。
  • 使用≥10个脉冲的中位数校正时,成核与塌缩均可达到>96%的水听器校正压力幅度。

诊断标准与关键数值

  • ACE校正平均获得水听器校正压力幅度的96%±3%,无校正时为71%±5%。
  • 使用≥10个脉冲的中位数时,成核与塌缩校正均获得>96%的水听器校正压力。
  • 单脉冲校正时,成核获得水听器校正压力的49%-99%,塌缩获得95%-99%。

临床应用与阅片要点

  • 在组织碎化术治疗中,可利用空化塌缩声发射进行快速像差校正,减少对水听器校正的依赖。
  • 单脉冲校正时优先选择塌缩信号,因其压力恢复范围更稳定(95%-99%)。
  • 若采用成核信号校正,建议使用≥10个脉冲的中位数以提高稳健性。

局限性

  • 研究为离体猪肝模型,未在活体或临床环境中验证。
  • 仅评估压力幅度恢复,未报告组织碎化术的实际疗效终点。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Phase aberration from soft tissue limits the efficacy of histotripsy, a therapeutic ultrasound technique based on acoustic cavitation. Previous work has shown that the acoustic emissions from cavitation can serve as "point sources" for aberration correction (AC). This study compared the efficacy of soft tissue AC for histotripsy using acoustic cavitation emissions (ACE) from bubble cloud nucleation and collapse.

METHODS: A 750-kHz, receive-capable histotripsy array was pulsed to generate cavitation in ex vivo porcine liver through an intervening abdominal wall. Received ACE signals were used to determine the arrival time differences to the focus and compute corrective delays. Corrections from single pulses and from the median of multiple pulses were tested.

DISCUSSION: On average, ACE AC obtained 96% ± 3% of the pressure amplitude obtained by hydrophone-based correction (compared with 71% ± 5% without AC). Both nucleation- and collapse-based corrections obtained >96% of the hydrophone-corrected pressure when using medians of ≥10 pulses. When using single-pulse corrections, nucleation obtained a range of 49%-99% of the hydrophone-corrected pressure, while collapse obtained 95%-99%.

CONCLUSION: The results suggest that (i) ACE AC can recover nearly all pressure amplitude lost owing to soft tissue aberration and that (ii) the collapse signal permits robust AC using a small number of pulses.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Yeats, Lu, Sukovich, et al. Soft Tissue Aberration Correction for Histotripsy Using Acoustic Emissions From Cavitation Cloud Nucleation and Collapse. Ultrasound Med Biol. 2023. PMID: 36759271. doi:10.1016/j.ultrasmedbio.2023.01.004

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。