比较用于确定聚焦超声稳定与惯性空化水平的体模和动物指标
本研究通过体模和动物实验评估聚焦超声诱导微泡空化时多种空化监测指标的一致性,发现系统噪声可能掩盖稳定与惯性空化的转换,物理稀疏化(PH-SP)可锐化亚谐波峰并提高亚谐波灵敏度32%。
要点
- 系统噪声可能大到掩盖稳定与惯性空化之间的转换,影响监测可靠性。
- 物理稀疏化(PH-SP)可锐化亚谐波峰形,体模和动物研究均可见。
- 优化PH-SP复值谱实部能量的指标使亚谐波灵敏度较标准指标提高32%。
核心结论
- 现有和提出的空化监测指标的初步研究表明,系统噪声可能大到掩盖稳定与惯性空化之间的转换;物理稀疏化(PH-SP)在体模和动物研究中均锐化亚谐波峰形,但动物研究中系统噪声升高和受试者间自然变异限制了指标随压力等效趋势的验证。
研究设计与样本
- 研究采用体模和动物实验,使用共聚焦水听器在10 ms聚焦超声脉冲期间以32 ns间隔采样,产生321,500个点,经离散傅里叶变换(DFT)获得高分辨率频谱进行分析。
诊断标准与关键数值
- 原文报告,优化PH-SP复值谱实部能量的指标使亚谐波灵敏度较标准指标提高32%,未报告敏感度、特异度、AUC或阈值。
临床应用与阅片要点
- 作者提出,硬件低通滤波器与物理稀疏化结合选择性去除频谱失真,可通过提高亚谐波灵敏度来优化空化监测指标;去除初始噪声瞬态可改善频谱信噪比(SNR)。
局限性
- 动物研究中系统噪声水平升高,结合受试者间的自然变异,限制了对指标随压力变化等效趋势的验证。
- 暂态长度从采样值中数字去除的微小变化会显著改变所有谐波相关聚焦超声信号的频谱带宽。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The opening of the blood-brain barrier (BBB) to allow therapeutic drug passage can be achieved by inducing microbubble cavitation using focused ultrasound (FUS). This approach can be monitored through analysis of the received signal to distinguish between stable cavitation associated with safe BBB opening and inertial cavitation associated with blood vessel damage. In this study, FUS phantom and animal studies were used to evaluate the experimental conditions that generate several cross-consistent metrics having the potential to be combined for the reliable, automatic control of cavitation levels.
METHODS: Typical metrics for cavitation monitoring involve observing changes in the spectrum generated by applying the discrete Fourier transform (DFT) to the time domain signal detected using a hydrophone during FUS. A confocal hydrophone was used to capture emissions during a 10 ms FUS burst, sampled at 32 ns intervals, to produce 321,500 points and a high-resolution spectrum when transformed. The FUS spectra were analyzed to show the impact that equipment-transients and well-known DFT-related distortions had on the metrics used for cavitation control. A new approach, physical sparsification (PH-SP), was introduced to sharpen FUS spectral peaks and minimize the effect of these distortions.
DISCUSSION: It was demonstrated that the general spectral signal-to-noise ratio (SNR) could be improved by removing the initial noisy phantom hydrophone signal transient. Minor changes in the transient length digitally removed from the sampled values significantly changed the spectral bandwidths of all the harmonically related FUS signals. We evaluated signal processing techniques to minimize the impact these DFT-related distortions on area-under-the-curve (AUC) metric calculations, and we identified the advantages of using PH-SP and proposed new metrics when characterizing FUS spectral properties. The results show many second, third and sub-harmonic metrics provide cross-consistent evidence of changes between stable and inertial cavitation levels. Removing the first harmonic signal component with a hardware low-pass filter allowed the hydrophone gain to be boosted without introducing distortion, leading to an improved analysis of the sub-harmonic signal orders of magnitude smaller in intensity. Metrics that optimized the energy in the real component of the complex-valued PH-SP spectra provided a 32% increase in the sub-harmonic sensitivity compared to standard metrics.
CONCLUSION: A preliminary investigation of existing and proposed metrics showed that system noise could be large enough to mask the transition between stable and inertial cavitation. Strong narrowing of sub-harmonic peak shapes on applying physical sparsification (PH-SP) were seen in both phantom and animal studies. However, validating equivalent trends of the metrics with pressure were limited by the increased system noise level in the animal study combined with the natural variability between subjects studied. The combined use of hardware low-pass filters and physical sparsification to selectively removing distortions in the spectrum allowed the optimization of metrics for cavitation monitoring by improving the sub-harmonic sensitivity.