肺超声帧率对肺毛细血管出血影响的实验研究

Frame Rate Exposimetry for Pulmonary Capillary Hemorrhage During Lung Ultrasound.
PubMed 文献 · J Ultrasound Med · 2023-02-01

该研究在大鼠模型中评估了肺超声帧率(fps)对肺毛细血管出血(PCH)的影响,使用Philips iE33和L15-7io探头在9 MHz下以34、61、118 fps进行B模式扫描。结果显示PCH面积随帧率增加而降低(19.5、9.6、7.5 mm²),低帧率时因更多高幅度脉冲暴露而增加PCH,MIOS为0.5时未见明显PCH。

要点

  • 低帧率肺超声可能增加肺毛细血管出血风险,高帧率相对安全。
  • TI值可能作为PCH的预测指标,低帧率时TI较高。
  • MIOS≤0.5时未观察到显著PCH,提示安全阈值参考。

关键影像表现/征象

  • PCH面积随帧率增加而降低:34 fps为19.5 mm²(MIOS=1.0,TI=0.8,4080幅图像),61 fps为9.6 mm²(MIOS=1.0,TI=0.5,7320幅图像),118 fps为7.5 mm²(MIOS=1.1,TI=0.4,14160幅图像)。
  • 34 fps(TI=0.5,MIOS=0.8)时PCH面积为10.7 mm²,与61和118 fps相比无显著差异,提示TI有一定预测价值。

诊断标准与关键数值

  • MIIS阈值分别为34 fps 0.42、61 fps 0.46、118 fps 0.49。
  • MIOS设置0.5对应MIIS值0.35-0.39时未发现显著PCH。

临床应用与阅片要点

  • 日常肺超声检查中应关注帧率设置,避免过低帧率增加出血风险。
  • 可参考TI值评估PCH风险,MIOS≤0.5可能相对安全。

局限性

  • 研究为大鼠模型,结果外推至临床需谨慎。
  • 单中心实验设计,样本量未报告。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Lung ultrasound (LUS) is a powerful and accessible clinical tool for pulmonary diagnosis, but risk of pulmonary capillary hemorrhage (PCH) presents a safety issue. The dependence of PCH in a rat model of LUS was evaluated for image frames-per-second (fps) and associated on-screen Mechanical Index (MIOS ) and Thermal Index (TI).

METHODS: A Philips iE33 machine with L15-7io probe was used to scan anesthetized rats in a warmed water bath. B mode was applied at 9 MHz with settings of 34, 61 and 118 fps. After 2 minutes of exposure at an MIOS setting, samples were obtained for assessment of PCH areas on the lung surface. Ultrasound parameters were measured to determine the in situ MIIS at the lung surface.

RESULTS: The PCH trend counter-intuitively decreased with increasing fps, with areas of 19.5 mm2 for 34 fps (MIOS = 1.0, TI = 0.8, 4080 images), 9.6 mm2 at 61 fps (MIOS = 1.0, TI = 0.5, 7320 images) and 7.5 mm2 at 118 fps (MIOS = 1.1, TI = 0.4, 14,160 images). The PCH was not significantly different for 34 fps (TI = 0.5, MIOS = 0.8) (10.7 mm2 ), compared to 61 and 118 fps, above, indicating some value for the TI as a predictive indicator of PCH. MIIS thresholds were 0.42, 0.46, and 0.49 for 34, 61 and 118 fps, respectively.

CONCLUSIONS: The increase in PCH at low fps was associated with delivering more relatively high amplitude grazing pulse exposures during slower image scans. No significant PCH was found for the MIOS setting of 0.5, corresponding to in MIIS values of 0.35-0.39.

来源:J Ultrasound Med,查看原文。版权归原作者所有。