第一代光子计数探测器CT细胞外容积标测实现心肌表征并与MRI对照
单中心前瞻性研究纳入29例成人,同日行心脏光子计数探测器CT与MRI检查,比较单能、双能PCD CT与MRI T1 mapping测得的整体及心中部细胞外容积。双能PCD CT与MRI相关性好至优(ICC 0.81–0.90),但高估ECV约2%;单能PCD CT低估约3%;双能较单能辐射剂量降低40%。
要点
- 双能PCD CT测ECV与MRI相关性好,但需注意约2%高估。
- 单能PCD CT测ECV较MRI低估约3%。
- 双能PCD CT辐射剂量较单能降低40%。
关键影像表现/征象
- 双能与单能PCD CT测得的ECV高度相关(r=0.91,P<.001)。
- 与MRI相比,双能PCD CT对心中部及整体ECV呈强相关(r=0.82和0.91,均P<.001),信度好至优(ICC 0.81和0.90),但高估ECV约2%。
- 单能PCD CT与MRI关系相似,但低估ECV约3%。
诊断标准与关键数值
- 双能PCD CT与MRI相比,心中部及整体ECV的相关系数分别为r=0.82和0.91(均P<.001),ICC分别为0.81和0.90。
- 双能PCD CT较MRI高估ECV约2%,单能PCD CT较MRI低估ECV约3%。
- 双能PCD CT的容积CT剂量指数为10.1 mGy,单能PCD CT为16.8 mGy(P<.001),双能较单能降低40%。
临床应用与阅片要点
- 双能PCD CT可作为心肌ECV定量的可选方法,与MRI相关性好,但报告时需考虑约2%的高估偏差。
- 单能PCD CT测ECV较MRI低估约3%,解读时需注意其系统性偏低。
- 双能模式在保持与单能模式高度相关(r=0.91)的同时辐射剂量降低40%,有利于临床推广。
局限性
- 单中心前瞻性研究,样本量较小(29例),结论外推需谨慎。
- 仅纳入第一代光子计数探测器CT系统,结果可能不适用于其他设备或代次。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Background Photon-counting detector (PCD) CT provides comprehensive spectral data with every acquisition, but studies evaluating myocardial extracellular volume (ECV) quantification with use of PCD CT compared with an MRI reference remain lacking. Purpose To compare ECV quantification for myocardial tissue characterization between a first-generation PCD CT system and cardiac MRI. Materials and Methods In this single-center prospective study, adults without contraindication to iodine-based contrast media underwent same-day cardiac PCD CT and MRI with native and postcontrast T1 mapping and late gadolinium enhancement for various clinical indications for cardiac MRI (the reference standard) between July 2021 and January 2022. Global and midventricular ECV were assessed with use of three methods: single-energy PCD CT, dual-energy PCD CT, and MRI T1 mapping. Quantitative comparisons among all techniques were performed. Correlation and reliability between different methods of ECV quantification were assessed with use of the Pearson correlation coefficient (r) and the intraclass correlation coefficient. Results The final sample included 29 study participants (mean age ± SD, 54 years ± 17; 15 men). There was a strong correlation of ECV between dual- and single-energy PCD CT (r = 0.91, P < .001). Radiation dose was 40% lower with dual-energy versus single-energy PCD CT (volume CT dose index, 10.1 mGy vs 16.8 mGy, respectively; P < .001). In comparison with MRI, dual-energy PCD CT showed strong correlation (r = 0.82 and 0.91, both P < .001) and good to excellent reliability (intraclass correlation coefficients, 0.81 and 0.90) for midventricular and global ECV quantification, but it overestimated ECV by approximately 2%. Single-energy PCD CT showed similar relationship with MRI but underestimated ECV by 3%. Conclusion Myocardial tissue characterization with photon-counting detector CT-based quantitative extracellular volume analysis showed a strong correlation to MRI. © RSNA, 2023 Supplemental material is available for this article.