商用肝纤维化弹性成像模体的变异性评估
该研究评估了两批次CIRS Model 039肝纤维化模体的机械属性变异性,每批次含四种硬度类型,每种8-10个模体,在10个不同角度进行剪切波弹性成像测量,计算群速度和相位速度曲线。采用多水平函数主成分分析分解相位速度曲线,以评估模体间和模体内变异性。结果显示各类型群速度标准差在批次1小于0.04 m/s、批次2小于0.02 m/s,大多数模体的相位速度曲线落在95%置信区间内。结论认为MFPCA是分析模体间和模体内变异性的有效工具,可在减少测试模体数量的情况下估计新批次的变异性。
要点
- 商用肝纤维化模体的群速度标准差在批次1<0.04 m/s、批次2<0.02 m/s。
- 大多数模体的相位速度曲线落在其类型95%置信区间内。
- MFPCA可有效分解模体间和模体内变异性。
核心结论
- MFPCA是分析相位速度曲线模体间和模体内变异性的有效工具。
- 在已知充分测试批次的变异性时,可用减少的模体数量估计新批次的变异性。
研究设计与样本
- 研究类型为模体实验研究,评估两批次CIRS Model 039模体。
- 每批次含四种硬度类型,每种类型8-10个模体。
- 对每个模体在10个不同角度进行剪切波弹性成像测量,计算群速度和相位速度曲线。
- 采用多水平函数主成分分析分解相位速度曲线,以水平1和水平2成分得分方差分别代表模体间和模体内变异性。
诊断标准与关键数值
- 批次1各类型模体群速度标准差均小于0.04 m/s。
- 批次2各类型模体群速度标准差均小于0.02 m/s。
- 两批次各类型中大多数单个模体的相位速度曲线落在该类型的95%置信区间内。
临床应用与阅片要点
- 在临床质量保证和研究中使用肝纤维化模体时,应了解模体间和模体内存在机械属性变异。
- 评估新批次模体变异性时,可参考已知充分测试批次的变异性以减少测试模体数量。
局限性
- 研究仅涉及两批次CIRS Model 039模体,结论外推需谨慎。
- 原文未报告样本量计算或统计效能分析。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Liver fibrosis has been found to increase the mechanical stiffness of the liver. To mimic different stages of liver fibrosis, commercially available phantoms (Model 039, CIRS, Inc.) have been produced for clinical quality assurance and research purposes. The purpose of this study was to investigate the mechanical property variability of the phantoms in two lots of CIRS Model 039 phantoms.
METHODS: Each lot consisted of phantoms of four stiffness types, and there were 8-10 phantoms of each type. Shear wave elastography measurements were conducted on each phantom at 10 different angles. Group velocity measurements and phase velocity curves were calculated for every SWE acquisition. Multilevel functional principal component analysis (MFPCA) was performed on phase velocity data, which decomposes each phase velocity curve into the sum of eigenfunctions of two levels. The variance of the component scores of levels 1 and 2 were used to represent inter-phantom and intra-phantom variability, respectively. The 95% confidence intervals of phase velocity in a phantom type were calculated to reflect curve variability.
DISCUSSION: The standard deviations of the group velocity for phantoms of any type were less than 0.04 and 0.02 m/s for lots 1 and 2, respectively. For both lots, in every type, the phase velocity curves of most individual phantoms fall within the 95% confidence interval.
CONCLUSION: MFPCA is an effective tool for analyzing the inter- and intra-phantom variability of phase velocity curves. Given the known variability of a fully tested lot, estimation of the variability of a new lot can be performed with a reduced number of phantoms tested.