文献速递心血管CT

光子计数探测器冠状动脉CTA:虚拟单能成像与迭代重建对图像质量的影响

Photon-counting detector coronary CT angiography: impact of virtual monoenergetic imaging and iterative reconstruction on image quality.
PubMed 文献 · Br J Radiol · 2023-01-12

研究评估光子计数探测器冠状动脉CTA中低keV虚拟单能图像与迭代重建对图像质量的影响,采用动态体模和10例患者,结果显示40 keV联合QIR-4可显著提升图像质量且不改变狭窄测量。

要点

  • 40 keV VMI联合QIR-4在体模中CNR、血管锐利度和衰减最高,患者主观评分最佳。
  • 狭窄测量不受keV或QIR水平影响,平均误差约3%/6%。
  • 噪声从70 keV QIR-4到40 keV QIR-off增加69%,但40 keV QIR-4噪声可接受。

核心结论

  • 40 keV虚拟单能图像联合QIR-4可显著改善光子计数探测器冠状动脉CTA的图像质量。

研究设计与样本

  • 前瞻性研究,使用高端心血管体模(含动态血流、搏动性心脏运动及不同钙化斑块狭窄分级)和10例连续患者。
  • PCD-CT扫描采用前瞻性ECG触发、120 kVp、自动管电流调制。
  • 重建40、50、60、70 keV虚拟单能图像,分别在不开启和开启QIR-1至QIR-4迭代重建下进行。

关键影像表现/征象

  • 体模中噪声纹理仅轻微受keV和QIR影响,噪声从70 keV QIR-4到40 keV QIR-off增加69%。
  • 40 keV QIR-4重建的CNR最高(46.1 ± 1.8),血管锐利度最高(425 ± 42 ∆HU/mm),血管衰减最高(1098 ± 14 HU)。
  • 患者主观评分中,40 keV QIR-3和QIR-4图像在所有主观类别和两位读者中得分最佳(p < 0.001)。

诊断标准与关键数值

  • 狭窄测量不受keV或QIR水平影响(p > 0.12),读者1和读者2的平均误差分别为3%和6%。
  • 患者图像主观评分采用5分Likert量表,评估整体图像质量、主观噪声、血管锐利度、血管对比度和冠状动脉斑块 delineation。

临床应用与阅片要点

  • 在光子计数探测器冠状动脉CTA中,可优先选择40 keV虚拟单能图像联合QIR-4重建以优化图像质量。
  • 狭窄测量在不同keV和QIR水平下稳定,阅片时无需担心重建参数影响狭窄评估。

局限性

  • 患者样本量较小(10例连续患者),可能限制统计效能和普遍性。
  • 体模结果可能不能完全反映体内复杂条件。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To assess the impact of low kilo-electronvolt (keV) virtual monoenergetic image (VMI) energies and iterative reconstruction on image quality of clinical photon-counting detector coronary CT angiography (CCTA).

METHODS: CCTA with PCD-CT (prospective ECG-triggering, 120 kVp, automatic tube current modulation) was performed in a high-end cardiovascular phantom with dynamic flow, pulsatile heart motion, and including different calcified plaques with various stenosis grades and in 10 consecutive patients. VMI at 40,50,60 and 70 keV were reconstructed without (QIR-off) and with all quantum iterative reconstruction (QIR) levels (QIR-1 to 4). In the phantom, noise power spectrum, vessel attenuation, contrast-to-noise-ratio (CNR), and vessel sharpness were measured. Two readers graded stenoses in the phantom and graded overall image quality, subjective noise, vessel sharpness, vascular contrast, and coronary artery plaque delineation on 5-point Likert scales in patients.

RESULTS: In the phantom, noise texture was only slightly affected by keV and QIR while noise increased by 69% from 70 keV QIR-4 to 40 keV QIR-off. Reconstructions at 40 keV QIR-4 exhibited the highest CNR (46.1 ± 1.8), vessel sharpness (425 ± 42 ∆HU/mm), and vessel attenuation (1098 ± 14 HU). Stenosis measurements were not affected by keV or QIR level (p > 0.12) with an average error of 3%/6% for reader 1/reader 2, respectively. In patients, across all subjective categories and both readers, 40 keV QIR-3 and QIR-4 images received the best scores (p < 0.001).

CONCLUSION: Forty keV VMI with QIR-4 significantly improved image quality of CCTA with PCD-CT.

ADVANCES IN KNOWLEDGE: PCD-CT at 40 keV and QIR-4 improves image quality of CCTA.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Sartoretti, McDermott, Mergen, et al. Photon-counting detector coronary CT angiography: impact of virtual monoenergetic imaging and iterative reconstruction on image quality. Br J Radiol. 2023. PMID: 36633005. doi:10.1259/bjr.20220466

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。