超声定量肝脂肪含量测量在呼吸时相间是否存在差异

Do Ultrasound Based Quantitative Hepatic Fat Content Measurements Have Differences Between Respiratory Phases?
PubMed 文献 · Acad Radiol · 2023-01-09

本研究纳入50例非酒精性脂肪性肝病(NAFLD)患者,使用组织衰减成像(TAI)和组织散射分布成像(TSI)分别测量衰减系数(AC)和散射分布系数(SDC),比较吸气峰值与呼气末的测量值。结果显示SDC在呼吸时相间无显著差异;AC在BMI≥30 kg/m²亚组无显著差异,但在BMI<30 kg/m²亚组呼吸时相间存在显著差异。

要点

  • SDC测量不受呼吸时相影响,可任意时相采集。
  • AC测量在BMI<30 kg/m²患者中需注意呼吸时相影响。
  • BMI≥30 kg/m²患者AC测量受呼吸时相影响不显著。

核心结论

  • 呼吸时相对SDC测量无影响,对AC测量在BMI<30 kg/m²亚组有显著影响,在BMI≥30 kg/m²亚组无显著影响。

研究设计与样本

  • 单中心研究,纳入50例NAFLD患者,使用TAI和TSI工具测量AC和SDC,每个呼吸时相各测量5次取中位数,采用Wilcoxon符号秩检验比较。

诊断标准与关键数值

  • 吸气峰值与呼气末AC中位数分别为0.87 dB/cm/MHz和0.89 dB/cm/MHz;SDC中位数分别为97.91和96.62。
  • 除BMI<30 kg/m²亚组的AC测量外,AC和SDC在呼吸时相间均无统计学显著差异。

临床应用与阅片要点

  • 测量SDC时无需控制呼吸时相;测量AC时,对BMI<30 kg/m²患者应统一呼吸时相以减少变异。

局限性

  • 单中心小样本研究,亚组分析样本量可能较小。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: The recently developed ultrasound based tools using attenuation coefficient (AC) and scatter distribution coefficient (SDC) values can be used to quantify hepatic fat content in patients with non-alcoholic fatty liver disease (NAFLD). However, currently the impact of respiratory phase on these measurements is not known. The purpose of this study is to compare AC and SDC measurements acquired at peak inspiration and end expiration phases.

MATERIALS AND METHODS: AC and SDC measurements were obtained in 50 patients with NAFLD. Tissue Attenuation Imaging (TAI) and Tissue Scatter Distribution Imaging (TSI) tools were utilized to measure AC and SDC values, respectively. Five measurements were performed at respiratory phases using TAI and TSI tools and the median values were noted. Subgroup analyses were performed and Wilcoxon signed rank test was used for comparison of the measurements.

RESULTS: The median values of the AC measurements at peak inspiration and end expiration phases were 0.87 dB/cm/MHz and 0.89 dB/cm/MHz, respectively. The median values of the SDC measurements at peak inspiration and end expiration phases were 97.91 and 96.62, respectively. There were no statistically significant differences in AC and SDC measurements between the respiratory phases except for AC measurements in BMI <30 kg/m2 subgroup.

CONCLUSION: Our results revealed that respiratory phases have no impact on SDC measurements. However, while the AC measurements in BMI ≥30 kg/m2 subgroup showed no significant difference, there was a significant difference in AC measurements in BMI <30 kg/m2 subgroup between the respiratory phases.