手持超声设备测量颈动脉直径的构建效度与可重复性

Construct Validity and Reproducibility of Handheld Ultrasound Devices in Carotid Artery Diameter Measurement.
PubMed 文献 · Ultrasound Med Biol · 2022-12-24

本研究评估三种常用手持超声设备测量颈动脉直径的构建效度和可重复性。体外实验显示各设备与高端超声的平均差异在0.009–0.023 cm之间,组内一致性均大于0.95;在体实验显示Telemed和Lumify观察者间一致性为优秀至良好,Butterfly为中等。结果表明手持超声设备可用于测量颈动脉直径。

要点

  • 手持超声测量颈动脉直径与高端超声一致性优秀,组内重复性均大于0.95。
  • Telemed和Lumify观察者间一致性优秀至良好,Butterfly仅为中等。
  • 手持超声可用于颈动脉直径测量,但Butterfly需注意观察者间变异。

核心结论

  • 手持超声设备测量颈动脉直径具有优秀的构建效度和组内可重复性,Telemed和Lumify观察者间一致性优秀至良好,Butterfly为中等。

研究设计与样本

  • 体外实验评估三种手持超声设备(Telemed MicrUs EXT-1H、Butterfly iQ、Philips Lumify)与高端超声对比的构建效度和组内变异。
  • 在体实验纳入20名健康志愿者,评估三种手持超声设备的观察者间变异。

诊断标准与关键数值

  • 与高端超声的平均差异:Telemed 0.023±0.030 cm,Butterfly 0.012±0.037 cm,Lumify 0.009±0.046 cm。
  • 与高端超声的组内一致性及手持设备组内变异均大于0.95,分类为优秀。
  • 观察者间一致性:Telemed 0.901(优秀),Lumify 0.827(良好),Butterfly 0.684(中等)。
  • 观察者间平均差异:Telemed 0.005±0.031 cm,Lumify 0.020±0.050 cm,Butterfly -0.003±0.033 cm。

临床应用与阅片要点

  • 手持超声可用于颈动脉直径测量,Telemed和Lumify观察者间一致性较好,适合多操作者场景。
  • Butterfly观察者间一致性中等,用于随访或多人测量时需注意变异。

局限性

  • 在体实验仅纳入20名健康志愿者,样本量小且未包括颈动脉病变患者。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

The construct validity and reproducibility of three commonly used handheld ultrasound (US) devices in measuring carotid arterial diameter was evaluated: Telemed MicrUs EXT-1H (Telemed, Vilnius, Lithuania), Butterfly iQ (Butterfly Network, Inc., Guilford, CT, USA) and Philips Lumify (Philips Healthcare, Best, The Netherlands). An in vitro setup was built to evaluate construct validity, compared with high-end US, and intra-observer variability of handheld US devices. Handheld devices showed a mean difference of 0.023 ± 0.030 cm, 0.012 ± 0.037 cm and 0.009 ± 0.046 cm for, respectively, Telemed, Butterfly and Lumify in comparison with high-end US devices. Intraclass agreement with the high-end system as well as intra-observer variability for handheld US devices was classified as excellent, with all values greater than 0.95. Subsequently, inter-observer variability of handheld US devices was investigated in an in vivo setup with 20 healthy volunteers. Inter-observer variability was classified as excellent for Telemed (0.901), good for Lumify (0.827) and moderate for Butterfly (0.684) with a difference of, respectively, 0.005 ± 0.031 cm, 0.020 ± 0.050 cm and -0.003 ± 0.033 cm. In conclusion, handheld US devices demonstrated an excellent construct validity and intra-observer variability. Additionally, excellent-to-good inter-observer variability for Telemed and Lumify was observed, and Butterfly demonstrated a moderate inter-observer agreement. These results indicate that handheld US devices are effective for measuring carotid arterial diameter.