压缩感知SPACE高分辨率3D T2WI在前列腺MRI中的诊断性能对比研究

High-resolution 3D T2-weighted SPACE sequence with compressed sensing for the prostate gland: diagnostic performance in comparison with conventional T2-weighted images.
PubMed 文献 · Abdom Radiol (NY) · 2022-12-22

该研究纳入179例前列腺穿刺前MRI患者,对比压缩感知SPACE 3D T2WI(层厚0.6 mm,扫描4分55秒)与常规TSE T2WI(层厚3 mm,扫描10分4秒)对临床显著癌检出(PI-RADS分类)和包膜外侵犯评估的诊断性能。结果显示两种序列PI-RADS分类AUC均高于0.7且无显著差异,EPE评估的敏感度和特异度亦无显著差异,CS SPACE在更短采集时间内达到与常规TSE相当的诊断性能。

要点

  • CS SPACE 3D T2WI层厚0.6 mm,扫描时间较三平面TSE T2WI缩短约一半。
  • 两种T2WI序列PI-RADS分类AUC均高于0.7,对临床显著癌检出无显著差异。
  • 两种序列评估包膜外侵犯的敏感度和特异度无显著差异。

核心结论

  • 压缩感知SPACE高分辨率3D T2WI在更短采集时间内,对临床显著癌检出和包膜外侵犯评估的诊断性能与常规TSE T2WI相当。

研究设计与样本

  • 单中心研究纳入179例前列腺穿刺前MRI患者(平均年龄69.1±9.3岁),行三平面TSE T2WI(层厚3 mm,扫描10分4秒)和CS SPACE(层厚0.6 mm,扫描4分55秒)。
  • 两名放射科医生独立评估两组图像,采用PI-RADS分类并判断病灶包膜外侵犯,以穿刺及前列腺切除标本为参考标准比较诊断性能。

关键影像表现/征象

  • CS SPACE序列层厚0.6 mm,较TSE T2WI的3 mm提供更薄层高分辨率图像。
  • 70例接受前列腺切除术的患者中25例确认包膜外侵犯。

诊断标准与关键数值

  • 两种T2WI序列PI-RADS分类的ROC曲线下面积均高于0.7,两名医生比较均无显著差异(p>0.05)。
  • 评估包膜外侵犯时,两种T2WI序列的敏感度和特异度无显著差异(p>0.05)。

临床应用与阅片要点

  • CS SPACE可在约5分钟内完成高分辨率3D T2WI采集,适合作为前列腺MRI T2WI的替代序列以缩短检查时间。
  • 阅片时注意CS SPACE层厚仅0.6 mm,需适应薄层图像上病灶形态和包膜外侵犯的判断。

局限性

  • 单中心研究,样本来自穿刺前MRI人群,包膜外侵犯评估仅基于70例接受前列腺切除术的患者。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To compare the diagnostic performance of high-resolution 3D T2-weighted imaging (T2WI) with compressed sensing (CS) sampling perfection with application-optimized contrasts using different flip angle evolution (SPACE) to that of conventional T2WI with turbo spin echo (TSE) in prostate MRI.

MATERIALS AND METHODS: This study evaluated 179 patients (mean age 69.1 ± 9.3) who underwent prostate biopsy after prostate prebiopsy MRI, including two sets of three-plane T2WI with TSE (thickness: 3 mm, scan time: 10 min 4 s) and CS SPACE (thickness: 0.6 mm, scan time: 4 min 55 s). Two radiologists evaluated two sets of images with the Prostate Imaging-Reporting and Data System (PIRADS) classification and determined the extraprostatic extension (EPE) of the lesion. The diagnostic performance to detect prostate cancer (PIRADS classification) and EPE was compared between the two T2WI sets.

RESULTS: Clinically significant cancer (CSC) was diagnosed in 103 patients (57.5%). Areas under the receiver operating characteristic curve of the PIRADS classification on both image sets with T2 TSE and T2 CS SPACE were higher than 0.7 and did not show significant differences for either radiologist (p > 0.05). EPE was confirmed in 25 of 70 patients underwent prostatectomy. For evaluating EPE on MRI, the sensitivity and specificity did not differ between the two T2WI sequences (p > 0.05).

CONCLUSION: High-resolution 3D T2WI using CS SPACE, which was acquired within a shorter acquisition time than three-plane T2 TSE, showed comparable diagnostic performance to conventional T2 TSE for detecting CSC and evaluating EPE.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。

引用

  1. Choi, Lee, Jung, et al. High-resolution 3D T2-weighted SPACE sequence with compressed sensing for the prostate gland: diagnostic performance in comparison with conventional T2-weighted images. Abdom Radiol (NY). 2022. PMID: 36544053. doi:10.1007/s00261-022-03777-x

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。