文献速递介入技术与安全DSA

介入放射学中医患剂量整合于剂量管理系统的实践

Occupational and patient doses for interventional radiology integrated into a dose management system.
PubMed 文献 · Br J Radiol · 2023-01-18

该研究将患者与职业剂量通过剂量管理系统整合管理,使用电子个人剂量计监测5个介入手术室的数据,2021年约4500例手术、8000条职业暴露记录,其中3043例有完整医患剂量数据(11名心脏科医生、7名放射科医生、8名护士)。系统可对患者剂量指标和职业暴露设置警报,实现单例手术的审计与优化。

要点

  • 剂量管理系统可整合患者与职业剂量,实现单例手术审计与警报设置。
  • 电子个人剂量计可区分不同辐射事件,弥补被动剂量计仅累积剂量的不足。
  • 通过审计患者剂量指标、职业剂量及比值,可优化辐射防护。

局限性

  • 原文未明确说明研究设计类型(如回顾性、前瞻性)及样本量计算依据。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The International Commission on Radiological Protection recommends managing patient and occupational doses as an integrated approach, for the optimisation of interventional procedures. The conventional passive personal dosimeters only allow one to know the accumulated occupational doses during a certain period of time. This information is not enough to identify if there is a lack of occupational radiation protection during some procedures. This paper describes the use of a dose management system (DMS) allowing patient and occupational doses for individual procedures to be audited.

METHODS: The DMS manages patient and occupational doses measured by electronic personal dosimeters. One dosemeter located at the C-arm is used as a reference for scatter radiation. Data have been collected from five interventional rooms. Dosimetry data can be managed for the whole procedure and the different radiation events. Optimisation is done through auditing different sets of parameters for individual procedures: patient dose indicators, occupational dose values, the ratio between occupational doses, and the doses measured by the reference dosemeter at the C-arm, and the ratio between occupational and patient dose values.

RESULTS: The managed data correspond to the year 2021, with around 4500 procedures, and 8000 records on occupational exposures. Patient and staff dose data (for 11 cardiologists, 7 radiologists and 8 nurses) were available for 3043 procedures. The DMS allows alerts for patient dose indicators and occupational exposures to be set.

CONCLUSIONS: The main advantage of this integrated approach is the capacity to improve radiation safety for patients and workers together, auditing alerts for individual procedures.

ADVANCES IN KNOWLEDGE: The management of patient and occupational doses together (measured with electronic personal dosimeters) for individual interventional procedures, using dose management systems, allows alerting optimisation on high-dose values for patients and staff.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Vano, Fernandez-Soto, Ten, et al. Occupational and patient doses for interventional radiology integrated into a dose management system. Br J Radiol. 2023. PMID: 36533561. doi:10.1259/bjr.20220607

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。