空间与时间分辨率对心血管磁共振特征追踪测量影响的数值体模研究
本研究基于一名志愿者的CMR图像构建了已知心肌运动的单层数值体模,在多种平面内空间分辨率和时间分辨率下,用两种特征追踪软件(MITT和两个版本的cvi42)测量应变并与真实值比较。结果显示不同厂商软件的特征追踪结果存在差异,且分辨率变化会影响应变测量的准确性,cvi42 v5.13.7的径向应变低估更明显。
要点
- 不同厂商及版本的特征追踪软件应变结果存在差异,报告时应注明软件及版本。
- 平面内空间分辨率降低可导致峰值应变与真实值的偏差增大。
- 数值体模可用于特征追踪软件的验证与比较,但临床推广前需进一步验证。
诊断标准与关键数值
- 以体模设定的真实径向和环向应变为参考标准,比较软件测量值与真实值的差异。
- 空间分辨率从1.4×1.4 mm²增至3.0×3.0 mm²时,cvi42 v5.13.7的峰值应变与真实值差异最大达9.17个百分点(径向)和8.42个百分点(环向)。
临床应用与阅片要点
- 日常阅片报告应变结果时应注明特征追踪软件名称及版本,避免跨厂商直接比较。
- 关注空间分辨率对 strain 测量的影响,低分辨率图像可能增大应变低估风险。
- 不同软件版本间的应变差异可能影响临床判读,升级软件后需重新评估测量一致性。
局限性
- 体模基于单名志愿者图像构建,可能不能完全代表不同患者的心脏运动特征。
- 仅测试了两种软件(MITT和cvi42两个版本),结论外推至其他软件需谨慎。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To develop a single-slice numerical phantom with known myocardial motion, at several temporal and in-plane spatial resolutions, for testing and comparison of Cardiovascular Magnetic Resonance (CMR) feature tracking (FT) software.
METHODS: The phantom was developed based on CMR acquisitions of one volunteer (acquired cine, tagging cine, T1 map, T2 map, proton density weighted image). The numerical MRI simulator JEMRIS was used, and the phantom was generated at several in-plane spatial resolutions (1.4 × 1.4 mm2 to 3.0 × 3.0 mm2) and temporal resolutions (20 to 40 cardiac phases). Two feature tracking software packages were tested: Medical Image Tracking Toolbox (MITT) and two versions of cvi42 (v5.3.8 and v5.13.7). The effect of resolution on strain results was investigated with reference to ground-truth radial and circumferential strain.
RESULTS: Peak radial strain was consistently undermeasured more for cvi42 v5.13.7 than for v5.3.8. Increased pixel size produced a trend of increased difference from ground-truth peak strain, with the largest changes for cvi42 obtained using v5.13.7 between 1.4 × 1.4 mm2 and 3.0 × 3.0 mm2, at 9.17 percentage points (radial) and 8.42 percentage points (circumferential).
CONCLUSIONS: The results corroborate the presence of intervendor differences in feature tracking results and show the magnitude of strain differences between software versions.
ADVANCES IN KNOWLEDGE: This study shows how temporal and in-plane spatial resolution can affect feature tracking with reference to the ground-truth strain of a numerical phantom. Results reaffirm the need for numerical phantom development for the validation and testing of FT software.