光子计数探测器CT低剂量胸部扫描对肺结节体积与气道壁厚度测量准确性的体模研究
该体模研究比较了高分辨率光子计数探测器CT(PCD-CT)与能量积分探测器CT(EID-CT)在低剂量胸部CT下对肺结节体积和气道壁厚度定量的准确性。结果显示,在0.1 mGy剂量下,超锐核PCD-CT对实性和磨玻璃结节体积误差(3.1%和3.3%)低于中锐核PCD-CT(9.9%和10.2%)和EID-CT(11.4%和9.2%);在0.2和0.8 mGy下,超锐核PCD-CT的气道壁厚度均方根误差(0.23 mm和0.16 mm)也低于EID-CT(0.42 mm和0.41 mm)及中锐核PCD-CT(0.54 mm和0.49 mm)。结论认为超锐核PCD-CT在更低辐射剂量下能提供更准确的结节体积和气道壁厚度定量。
要点
- 超锐核PCD-CT在0.1 mGy低剂量下肺结节体积测量误差最小,优于EID-CT和中锐核PCD-CT。
- 超锐核PCD-CT在0.2和0.8 mGy下气道壁厚度测量误差显著低于EID-CT和中锐核PCD-CT。
- 中锐核PCD-CT图像噪声最低,但超锐核PCD-CT在定量准确性上更优。
诊断标准与关键数值
- 在0.1 mGy下,实性结节体积平均误差:超锐核PCD-CT 3.1%,中锐核PCD-CT 9.9%,EID-CT 11.4%。
- 在0.1 mGy下,磨玻璃结节体积平均误差:超锐核PCD-CT 3.3%,中锐核PCD-CT 10.2%,EID-CT 9.2%。
- 在0.2和0.8 mGy下,气道壁厚度均方根误差:EID-CT 0.42 mm和0.41 mm,中锐核PCD-CT 0.54 mm和0.49 mm,超锐核PCD-CT 0.23 mm和0.16 mm。
局限性
- 本研究为体模研究,结果需在临床患者中进一步验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: A comparison of high-resolution photon-counting detector computed tomography (PCD-CT) versus energy-integrating detector (EID) CT via a phantom study using low-dose chest CT to evaluate nodule volume and airway wall thickness quantification.
MATERIALS AND METHODS: Twelve solid and ground-glass lung nodule phantoms with 3 diameters (5 mm, 8 mm, and 10 mm) and 2 shapes (spherical and star-shaped) and 12 airway tube phantoms (wall thicknesses, 0.27-1.54 mm) were placed in an anthropomorphic chest phantom. The phantom was scanned with EID-CT and PCD-CT at 5 dose levels (CTDI vol = 0.1-0.8 mGy at Sn-100 kV, 7.35 mGy at 120 kV). All images were iteratively reconstructed using matched kernels for EID-CT and medium-sharp kernel (MK) PCD-CT and an ultra-sharp kernel (USK) PCD-CT kernel, and image noise at each dose level was quantified. Nodule volumes were measured using semiautomated segmentation software, and the accuracy was expressed as the percentage error between segmented and reference volumes. Airway wall thicknesses were measured, and the root-mean-square error across all tubes was evaluated.
RESULTS: MK PCD-CT images had the lowest noise. At 0.1 mGy, the mean volume accuracy for the solid and ground-glass nodules was improved in USK PCD-CT (3.1% and 3.3% error) compared with MK PCD-CT (9.9% and 10.2% error) and EID-CT images (11.4% and 9.2% error), respectively. At 0.2 mGy and 0.8 mGy, the wall thickness root-mean-square error values were 0.42 mm and 0.41 mm for EID-CT, 0.54 mm and 0.49 mm for MK PCD-CT, and 0.23 mm and 0.16 mm for USK PCD-CT.
CONCLUSIONS: USK PCD-CT provided more accurate lung nodule volume and airway wall thickness quantification at lower radiation dose compared with MK PCD-CT and EID-CT.
引用
- Dunning, Marsh, Winfree, et al. Accuracy of Nodule Volume and Airway Wall Thickness Measurement Using Low-Dose Chest CT on a Photon-Counting Detector CT Scanner. Invest Radiol. 2022. PMID: 36525385. doi:10.1097/RLI.0000000000000933
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。