3T场下胎儿全脊柱加速三维磁敏感加权成像研究
研究纳入37例超声怀疑胎儿椎体异常的孕妇,在3T下行3D-SWI、3D True FISP和2D HASTE扫描,比较采集时间、SNR、CNR及诊断效能。结果显示3D-SWI采集时间短、CNR最高,对椎体异常诊断AUC达0.92,可作为常规MR的补充方法。
要点
- 3D-SWI采集仅15秒,较HASTE的37秒显著缩短,适合胎儿快速成像。
- 3D-SWI的CNR最高,对颈段和骶尾段椎体显示优于HASTE。
- 3D-SWI诊断胎儿椎体异常的AUC为0.92,诊断效能最高。
核心结论
- 3D-SWI可在3T场下以较短采集时间实现胎儿全脊柱的清晰显示,对椎体畸形诊断准确性高,可作为常规MR的补充序列。
研究设计与样本
- 单中心前瞻性研究,纳入37例孕22-39周(平均29±3周)超声怀疑胎儿椎体异常的孕妇,在3.0T MR上行3D-SWI、2D HASTE和3D True FISP扫描,记录采集时间并比较图像质量与诊断效能。
关键影像表现/征象
- 3D-SWI的CNR最高,3D True FISP的SNR最高。
- 3D-SWI与3D True FISP对颈段和骶尾段椎体的显示优于HASTE。
- 3D-SWI对全脊柱各段显示与3D True FISP相当。
- 图像质量主观评分观察者间加权Kappa值为0.70-0.89,一致性良好至优秀。
诊断标准与关键数值
- 3D-SWI诊断胎儿椎体异常的AUC为0.92,为三种序列中最高。
- 3D-SWI采集时间15秒,3D True FISP为17秒,均显著短于HASTE的37秒(均P<0.01)。
临床应用与阅片要点
- 在超声怀疑胎儿椎体异常时,可加做3D-SWI序列快速评估全脊柱,尤其关注颈段和骶尾段。
- 3D-SWI的CNR优势有助于显示椎体畸形,但SNR低于3D True FISP,需结合其他序列综合判断。
局限性
- 单中心小样本研究,仅纳入超声怀疑椎体异常的胎儿,可能存在选择偏倚。
- 未报告组织学或出生后随访确诊结果作为参考标准。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To investigate the image quality and capability of generalized auto-calibrating partially parallel acquisition (GRAPPA) accelerated Three-dimensional (3D) susceptibility weighted imaging (SWI) of the whole spine at 3T.
METHODS: A total of 37 pregnant women (gestation age 22 to 39 weeks, average 29 ± 3 weeks) with suspected fetal vertebral anomalies by ultrasound (US) screening underwent 3.0T MR imaging with 3D SWI, conventional two-dimensional (2D) half-flourier acquisition single-shot turbo spin-echo (HASTE) and 3D true fast imaging with steady-state precession (True FISP). The acquisition time of each protocol was recorded. Signal-to-noise ratios (SNRs) and contrast-to-noise ratios (CNRs) were determined in representative interest regions of fetal thoracic vertebrae and compared among three pulse sequences. Two radiologists rated image quality independently in random order on a 5-point scale. Kappa coefficients were computed to assess inter-observer reliability. Receiver operating characteristic curves were generated, and the area under the curve (AUC) was used to compare the diagnostic performance of each protocol in vertebral deformities.
RESULTS: The acquisition time was 15 s for 3D-SWI and 17 s for 3D True FISP, significantly shorter than conventional HASTE (37 s; both P < 0.01). Of the three protocols, The SNR was highest on 3D True FISP, while the CNR was highest on 3D SWI. Visualization of all segments of the whole spine by 3D SWI was comparable with 3D True FISP. In contrast, 3D SWI and 3D True FISP depicted cervical and sacrococcygeal vertebrae better than HASTE. The weighted kappa statistic was 0.70-0.89 to evaluate the image quality of all segments of the whole spine, indicating good to excellent interobserver agreement. 3D SWI had the highest diagnostic performance for detecting fetal vertebral anomalies (AUC = 0.92).
CONCLUSIONS: 3D-SWI is feasible for improved visualization of the whole fetal vertebral column and its congenital malformations with adequate image quality and high accuracy, thereby providing a supplementary method to conventional MR imaging.
引用
- Liu, Li, Pang, et al. Accelerated three-dimensional susceptibility weighted imaging of the whole spine of fetus at 3T. Eur J Radiol. 2022. PMID: 36481479. doi:10.1016/j.ejrad.2022.110622
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。