双源光子计数探测器CT支架成像:管腔内衰减与锐利重建核对于管腔可见度的影响
该体模研究在双源光子计数探测器CT上评估扫描模式与重建核对于不同尺寸支架管腔可见度与锐利度的影响。结果显示,超高清模式(0.2 mm层厚)与锐利重建核(Qr72c、Bv60m)显著提高支架管腔可见度与锐利度,尤其对于小尺寸支架;管腔内衰减每增加200 HU,可见度下降2.1%。
要点
- 超高清模式(0.2 mm)联合锐利核(Qr72c或Bv60m)可显著改善支架管腔可见度,小支架获益最大。
- 管腔内对比剂浓度增高会降低支架管腔可见度,每200 HU下降2.1%。
- 评估小支架(如2.5 mm)时应优先选择超高清模式或Bv60m虚拟单能重建。
关键影像表现/征象
- 支架管腔可见度在Qr72c(0.2 mm)重建中最高(86.5±2.8%至88.3±2.6%),其次为Bv60m(0.4 mm)(77.3±2.9%至82.7±2.5%)。
- 最小支架(2.5 mm)管腔可见度最低(Qr40c(0.6 mm)为54.1%,Qr72c(0.2 mm)为74.1%),最大支架(9 mm)最高(Qr40c(0.6 mm)为93.8%,Qr72c(0.2 mm)为99.1%)。
- 支架与管腔最大ΔHU在Qr72c(0.2 mm)(ΔHU 892±504至1526±517)和Bv60m(0.4 mm)(ΔHU 480±357至1030±344)序列中最高。
- 超高清模式及Bv60m(0.4 mm)序列对小尺寸支架的管腔可见度和锐利度改善最明显。
诊断标准与关键数值
- 管腔可见度定义为支架管腔可见百分比,管腔内衰减每增加200 HU,管腔可见度下降2.1%。
临床应用与阅片要点
- 日常阅片评估支架通畅性时,建议采用超高清模式(0.2 mm)联合锐利核(Qr72c)或Bv60m虚拟单能重建,尤其针对小支架。
- 注意管腔内高对比剂浓度会降低支架管腔可见度,报告时应考虑此因素。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To assess the impact of scan modes and reconstruction kernels using a novel dual-source photon-counting detector CT (PCD-CT) on lumen visibility and sharpness of different stent sizes.
METHODS: A phantom containing six balloon-expandable stents (2.5 to 9 mm diameter) in silicone tubing was scanned on a PCD-CT with standard (0.6 mm and 0.4 mm thicknesses) and ultra-high-resolution (0.2 mm thickness) modes. With the use of increasing contrast medium concentrations, densities of 0, 200, 400, and 600 HU were achieved. Standard-resolution scans were reconstructed using increasing sharpness kernels, using both polyenergetic quantitative soft tissue "conventional" ((Qr40c(0.6 mm), Qr40c(0.4 mm), Qr72c(0.2 mm)) and vascular (Bv) virtual monoenergetic reconstructions (Bv44m(0.4 mm), Bv60m(0.4 mm)) at 70 keV. In-stent lumen visibility, sharpness (max. ΔHU of the stent measured in profile plots), and in-stent noise (standard deviation of HU) were measured.
RESULTS: In-stent lumen visibility was highest for Qr72c(0.2 mm) (86.5 ± 2.8% to 88.3 ± 2.6%) and in Bv60m(0.4 mm) reconstructions (77.3 ± 2.9 to 82.7 ± 2.5%). Lumen visibility was lowest in the smallest stent (2.5 mm) ranging from 54.1% in Qr40c(0.6 mm) to 74.1% in Qr72c(0.2 mm) and highest in the largest stent (9 mm) ranging from 93.8% in Qr40c(0.6 mm) to 99.1% in the Qr72c(0.2 mm) series. Lumen visibility decreased by 2.1% for every 200-HU increase in lumen attenuation. Max. ΔHU between stents and stent lumen was highest in Qr72c(0.2 mm) (ΔHU 892 ± 504 to 1526 ± 517) and Bv60m(0.4 mm) series (ΔHU 480 ± 357 to 1030 ± 344). Improvement of lumen visibility and sharpness in UHR and Bv60m(0.4 mm) series was strongest in smaller stent sizes.
CONCLUSION: UHR acquisition mode and sharp reconstruction kernels on a novel PCD-CT system significantly improve in-stent lumen visibility and sharpness-especially for smaller stent sizes.
KEY POINTS: • In-stent lumen visibility and sharpness of stents significantly improve using sharp reconstruction kernels (Bv60) and ultra-high-resolution mode in photon-counting detector computed tomography. • The observed improvement of stent-lumen visibility was highest in smaller stent sizes.
引用
- Decker, O'Doherty, Schoepf, et al. Stent imaging on a clinical dual-source photon-counting detector CT system-impact of luminal attenuation and sharp kernels on lumen visibility. Eur Radiol. 2022. PMID: 36462045. doi:10.1007/s00330-022-09283-4