高加速Wave-CAIPI增强3D-T1 MPRAGE与标准序列检测颅内强化病灶的验证对比
本研究纳入80例增强脑MRI患者,对比高加速Wave-CAIPI增强3D-T1 MPRAGE(R=4,采集时间2分32秒)与标准3D-T1 MPRAGE(R=2,采集时间5分18秒)。两名神经放射科医师评估显示,Wave序列在显示强化病灶方面非劣于标准序列,噪声、运动伪影和整体诊断质量感知相当,采集时间缩短一半以上。
要点
- Wave-T1-MPRAGE将3D增强T1采集时间从5分18秒缩短至2分32秒,图像质量非劣于标准序列。
- 两序列对颅内强化病灶显示一致率达100%(80/80例),无临床重要信息丢失。
- Wave序列在94%(75/80)病例中噪声感知与标准序列相当,16%(13/80)病例因运动伪影减少而更受偏好。
关键影像表现/征象
- Wave-T1-MPRAGE在显示颅内强化病灶方面与标准T1-MPRAGE非劣,所有病例两序列显示一致。
- Wave-T1-MPRAGE在噪声感知、运动伪影和整体诊断质量方面非劣于标准序列(p<0.001)。
- Wave序列在94%(75/80)病例中噪声感知与标准序列相当,在16%(13/80)病例中因运动伪影减少而更受偏好。
诊断标准与关键数值
- 非劣效性检验采用15%非劣效性界值,Wave-T1-MPRAGE在显示强化病灶、噪声感知、运动伪影和整体诊断质量方面均非劣于标准T1-MPRAGE(p<0.001)。
- 两序列对颅内强化病灶显示的一致率为100%(80/80例),观察者间和观察者内一致性均 unanimous。
临床应用与阅片要点
- Wave-T1-MPRAGE可显著缩短增强3D-T1采集时间,适合无法长时间配合或易产生运动伪影的患者。
- 在常规阅片中,Wave序列图像可用于评估颅内强化病灶,但需注意其噪声和伪影感知与标准序列相当,无临床重要信息丢失。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: High-resolution post-contrast T1-weighted imaging is a workhorse sequence in the evaluation of neurological disorders. The T1-MPRAGE sequence has been widely adopted for the visualization of enhancing pathology in the brain. However, this three-dimensional (3D) acquisition is lengthy and prone to motion artifact, which often compromises diagnostic quality. The goal of this study was to compare a highly accelerated wave-controlled aliasing in parallel imaging (CAIPI) post-contrast 3D T1-MPRAGE sequence (Wave-T1-MPRAGE) with the standard 3D T1-MPRAGE sequence for visualizing enhancing lesions in brain imaging at 3 T.
METHODS: This study included 80 patients undergoing contrast-enhanced brain MRI. The participants were scanned with a standard post-contrast T1-MPRAGE sequence (acceleration factor [R] = 2 using GRAPPA parallel imaging technique, acquisition time [TA] = 5 min 18 s) and a prototype post-contrast Wave-T1-MPRAGE sequence (R = 4, TA = 2 min 32 s). Two neuroradiologists performed a head-to-head evaluation of both sequences and rated the visualization of enhancement, sharpness, noise, motion artifacts, and overall diagnostic quality. A 15% noninferiority margin was used to test whether post-contrast Wave-T1-MPRAGE was noninferior to standard T1-MPRAGE. Inter-rater and intra-rater agreement were calculated. Quantitative assessment of CNR/SNR was performed.
RESULTS: Wave-T1-MPRAGE was noninferior to standard T1-MPRAGE for delineating enhancing lesions with unanimous agreement in all cases between raters. Wave-T1-MPRAGE was noninferior in the perception of noise (p < 0.001), motion artifact (p < 0.001), and overall diagnostic quality (p < 0.001).
CONCLUSION: High-accelerated post-contrast Wave-T1-MPRAGE enabled a two-fold reduction in acquisition time compared to the standard sequence with comparable performance for visualization of enhancing pathology and equivalent perception of noise, motion artifacts and overall diagnostic quality without loss of clinically important information.
KEY POINTS: • Post-contrast wave-controlled aliasing in parallel imaging (CAIPI) T1-MPRAGE accelerated the acquisition of three-dimensional (3D) high-resolution post-contrast images by more than two-fold. • Post-contrast Wave-T1-MPRAGE was noninferior to standard T1-MPRAGE with unanimous agreement between reviewers (100% in 80 cases) for the visualization of intracranial enhancing lesions. • Wave-T1-MPRAGE was equivalent to the standard sequence in the perception of noise in 94% (75 of 80) of cases and was preferred in 16% (13 of 80) of cases for decreased motion artifact.
引用
- Goncalves Filho, Awan, Conklin, et al. Validation of a highly accelerated post-contrast wave-controlled aliasing in parallel imaging (CAIPI) 3D-T1 MPRAGE compared to standard 3D-T1 MPRAGE for detection of intracranial enhancing lesions on 3-T MRI. Eur Radiol. 2022. PMID: 36460923. doi:10.1007/s00330-022-09265-6
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。