文献速递腹部介入US

环形换能器无创高强度聚焦超声治疗猪肝组织:快速、大范围且安全的消融

Non-invasive High-Intensity Focused Ultrasound Treatment of Liver Tissues in an In Vivo Porcine Model: Fast, Large and Safe Ablations Using a Toroidal Transducer.
PubMed 文献 · Ultrasound Med Biol · 2022-10-28

本研究在8只活体猪中,使用工作频率2.5MHz、32个等面积同心环的环形HIFU换能器,经皮徒手于呼吸暂停期进行肝脏消融,每只动物40秒内创建两个热消融灶。消融灶长轴20.9±6.3mm、短轴14.2±5.5mm,所有消融灶均可在超声显示,超声与大体病理尺寸相关系数r=0.84。提示该手持设备可快速形成大范围消融灶并实现可靠超声引导,可能成为肝肿瘤治疗新选择。

要点

  • 环形HIFU换能器可在40秒内于猪肝形成长轴约20.9mm、短轴约14.2mm的热消融灶。
  • 所有消融灶均可在超声显示,超声测值与大体病理尺寸相关性r=0.84。
  • 术前测量皮肤、脂肪、肌肉和肝脏衰减以调整聚焦区能量,腹壁平均厚度14.8±1.3mm。

关键影像表现/征象

  • 所有HIFU消融灶均可在超声声像图上显示。
  • 消融灶长轴为20.9±6.3mm(范围14.0-33.7mm),短轴为14.2±5.5mm(范围7.0-22.0mm)。
  • 超声测量的消融灶尺寸与大体病理测量尺寸的相关系数r=0.84。

诊断标准与关键数值

  • 消融灶尺寸以长轴和短轴表示,长轴20.9±6.3mm、短轴14.2±5.5mm。
  • 超声与大体病理尺寸相关性r=0.84。

临床应用与阅片要点

  • 术前需测量皮肤、脂肪、肌肉和肝脏组织的超声衰减,以调整送达聚焦区的能量。
  • 超声可显示HIFU消融灶,且测值与大体病理尺寸相关性良好,可作为术中引导和评估工具。

局限性

  • 研究为活体猪模型,样本量8只,结果向人体临床转化的有效性尚需验证。
  • 设备由手持固定,操作者依赖性可能影响消融灶的重复性和尺寸控制。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

A toroidal high-intensity focused ultrasound (HIFU) transducer was used to non-invasively treat liver tissues in vivo in a pig model. The transducer was divided into 32 concentric rings with equal surface areas operating at 2.5 MHz. First, attenuation of skin, fat, muscle and liver tissues was measured in fresh animal samples to adjust the energy delivered to the focal zone. Then, 8 animals were included in the present protocol and placed in a dorsal decubitus proclive position at an angle of 15°. The device was held by hand, and sonications were performed during apnea. Two thermal HIFU lesions were created in 40 s in each animal. The average abdominal wall thickness was 14.8 ± 1.3 mm (12.5-17.6 mm). The longest and shortest axes of the HIFU ablations were 20.9 ± 6.3 mm (14.0-33.7 mm) and 14.2 ± 5.5 mm (7.0-22.0 mm), respectively. All HIFU lesions were visible on sonograms. The correlation between the dimensions of the HIFU lesions observed on sonograms and those obtained during gross examination was r = 0.84. Creating large and fast ablations with reliable ultrasound imaging guidance in the liver using this handheld device may represent a new therapeutic option for patients with liver tumors.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Cambronero, Dupré, Mastier, et al. Non-invasive High-Intensity Focused Ultrasound Treatment of Liver Tissues in an In Vivo Porcine Model: Fast, Large and Safe Ablations Using a Toroidal Transducer. Ultrasound Med Biol. 2022. PMID: 36441030. doi:10.1016/j.ultrasmedbio.2022.08.015

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。