基于回波相位分割的优化孔径模式对比增强超声
该研究评估了不同幅度调制(AM)孔径模式对超声造影组织抑制和对比噪声比(CTR)的影响,并引入基于回波相位差的分割技术。结果表明,采用两个阵元交替的孔径模式(AM2)结合基于最大相位差(ΔΦmax)的分割,在流动体模和灌注猪肝中分别比传统AM联合ΔΦAM分割的CTR提高7 dB和6 dB,为最优发射孔径设计。
要点
- AM2孔径模式(两个阵元交替)比传统棋盘格孔径获得更高CTR。
- 基于回波相位差的分割技术可使所有孔径模式的CTR提高约25 dB。
- AM2联合ΔΦmax分割在体模和猪肝中均优于传统方法,为最优设计。
关键影像表现/征象
- AM2孔径模式(两个阵元交替)在评估的探头中具有最高的CTR。
- 基于回波相位差(ΔΦAM、ΔΦhalf、ΔΦmax)的分割技术可使所有孔径模式的CTR提高约25 dB。
- AM2联合ΔΦmax分割在流动体模中CTR比传统AM联合ΔΦAM分割高7 dB,在灌注猪肝中对比度高6 dB。
临床应用与阅片要点
- 在超声造影中可考虑使用AM2孔径模式替代传统棋盘格孔径,以提高对比噪声比。
- 应用基于回波相位差的分割技术(尤其是ΔΦmax)可进一步抑制组织信号,改善微泡检测。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Amplitude modulation (AM) suppresses tissue signals and detects microbubble signals in contrast-enhanced ultrasound (CEUS) and is often implemented with checkerboard apertures. However, possible crosstalk between transmitting and non-transmitting array elements may compromise tissue suppression in AM. Using AM aperture patterns other than the conventional checkerboard approach (one on, one off) may reduce the degree of crosstalk and increase the contrast-to-tissue-ratio (CTR) compared with conventional AM. Furthermore, previous studies have reported that the phase difference between the echoes in AM pulsing sequences may be used to segment tissue and microbubbles and improve tissue signal suppression and the CTR of CEUS images. However, the CTR of the image produced by alternative AM aperture patterns and the effect of segmentation approach on these alternative apertures have not been investigated. We evaluated a number of AM aperture patterns to find an optimal AM aperture pattern that provides the highest CTR. We found that the aperture that uses alternating groups of two elements, AM2, had the highest CTR for the probe evaluated. In addition, a segmentation technique based on echo phase differences (between the full and half-pulses, ΔΦAM, between the complementary half-pulses, ΔΦhalf, and the maximum of the two ΔΦmax) was also considered in the AM aperture optimization process. The segmentation approach increases the CTR by about 25 dB for all apertures. Finally, AM2 segmented with ΔΦmax had a 7-dB higher CTR in a flow phantom and a 6-dB higher contrast in a perfused pig liver than conventional AM segmented with ΔΦAM, and it is the optimal transmit aperture design.
引用
- Lai, Averkiou. Contrast-Enhanced Ultrasound with Optimized Aperture Patterns and Bubble Segmentation Based on Echo Phase. Ultrasound Med Biol. 2022. PMID: 36441029. doi:10.1016/j.ultrasmedbio.2022.08.013
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。