深度学习自动检测TOF-MRA脑动脉瘤的外部验证研究
软件总体敏感度72.6%、特异度87.2%、准确率82.6%,对大于5mm无血栓囊状动脉瘤敏感度达100%,但对梭形或血栓化动脉瘤检测率低。结论认为该软件可辅助放射科医生报告TOF-MRA,但梭形和血栓化动脉瘤需进一步改进。
要点
- AI软件对大于5mm无血栓囊状动脉瘤敏感度100%,阅片时可优先关注此类病变。
- 梭形动脉瘤检测率仅33.3%、血栓化动脉瘤仅16.7%,报告中需警惕漏诊。
- 软件总体敏感度72.6%、特异度87.2%,阴性预测值88.5%,可作为辅助筛查工具。
核心结论
- 该AI软件可辅助放射科医生报告TOF-MRA,对囊状动脉瘤检测高度可靠,但梭形或血栓化动脉瘤需进一步改进。
关键影像表现/征象
- 大于5mm无血栓囊状动脉瘤敏感度100%,梭形动脉瘤检测率33.3%,血栓化动脉瘤检测率16.7%。
- 8例初始书面报告未包含但专家回顾检出的动脉瘤中,软件检出4例。
诊断标准与关键数值
- 以经验丰富放射科医生读片为参考标准,软件检测颅内动脉瘤的敏感度72.6%、特异度87.2%、阳性预测值67.9%、阴性预测值88.5%、准确率82.6%。
- 按动脉瘤特征分层:大于5mm无血栓囊状动脉瘤敏感度100%,梭形动脉瘤33.3%,血栓化动脉瘤16.7%。
临床应用与阅片要点
- 阅片时可利用软件辅助筛查囊状动脉瘤,尤其关注大于5mm无血栓病变。
- 对梭形或血栓化动脉瘤不应依赖软件,需人工仔细评估避免漏诊。
- 软件阴性预测值88.5%,阴性结果不能完全排除动脉瘤,仍需结合原始图像判读。
局限性
- 原文指出需进一步研究评估软件对检出率、阅片者间一致性和阅读时间的影响。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Cerebral aneurysms yield the risk of rupture, severe disability and death. Thus, early detection of cerebral aneurysms is crucial to ensure timely treatment, if necessary. AI-based software tools are expected to enhance radiologists' performance in detecting pathologies like cerebral aneurysms in the future. Our aim was to evaluate the diagnostic performance of an artificial intelligence-based software designed to detect intracranial aneurysms on TOF-MRA.
MATERIALS AND METHODS: One hundred ninety-one MR imaging data sets were analyzed using the software mdbrain for the presence of intracranial aneurysms on TOF-MRA obtained using two 3T MR imaging scanners or a 1.5T MR imaging scanner according to our clinical standard protocol. The results were compared with the reading of an experienced radiologist as a criterion standard to measure the sensitivity, specificity, positive and negative predictive values, and accuracy of the software. Additionally, detection rates depending on size, morphology, and location of the aneurysms were evaluated.
RESULTS: Fifty-four aneurysms were detected by the expert reader. The overall sensitivity of the software for the detection of cerebral aneurysms was 72.6%, the specificity was 87.2%, and the accuracy was 82.6%. The positive predictive value was 67.9%, and the negative predictive value was 88.5%. We observed a sensitivity of 100% for saccular aneurysms of >5 mm without signs of thrombosis and low detection rates for fusiform or thrombosed aneurysms of 33.3% and 16.7%, respectively. Of 8 aneurysms that were not included in the initial written reports but were detected by the expert reader, retrospectively, 4 were detected by the software.
CONCLUSIONS: Our data suggest that the software can assist radiologists in reporting TOF-MRA. The software was highly reliable in detecting saccular aneurysms, while for fusiform or thrombosed aneurysms, further improvements are needed. Further studies are necessary to investigate the impact of the software on detection rates, interrater reliability, and reading times.
引用
- Lehnen, Haase, Schmeel, et al. Automated Detection of Cerebral Aneurysms on TOF-MRA Using a Deep Learning Approach: An External Validation Study. AJNR Am J Neuroradiol. 2022. PMID: 36357154. doi:10.3174/ajnr.A7695
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。