相干性指标实现乳腺囊实性肿块的阅片者无关鉴别
本研究基于31个乳腺肿块超声数据,提出滞后一相干(LOC)和相干长度(CL)两个客观指标,无需放射科医生判读即可定量区分囊性与实性肿块。
要点
- LOC均值或中位数可完美区分囊性与实性乳腺肿块(敏感度、特异度均1.00)。
- CL指标区分效能略低,均值CL敏感度/特异度1.00/0.95,中位数CL为0.92/0.89。
- 致密乳腺组织中鉴别敏感度和特异度最高,优于非致密组织。
关键影像表现/征象
- 基于每个肿块内LOC均值或中位数可完全区分液性填充与实性乳腺肿块。
- 基于CL均值区分时敏感度/特异度为1.00/0.95,基于CL中位数时为0.92/0.89。
- 致密乳腺组织中鉴别敏感度和特异度最高,优于非致密组织。
诊断标准与关键数值
- LOC均值或中位数区分囊实性肿块的敏感度和特异度均为1.00。
- CL均值区分囊实性肿块的敏感度/特异度为1.00/0.95。
- CL中位数区分囊实性肿块的敏感度/特异度为0.92/0.89。
临床应用与阅片要点
- 可利用LOC均值或中位数作为客观指标辅助鉴别乳腺囊实性肿块,减少阅片者间差异。
- 在致密乳腺组织中相干性指标鉴别效能更高,可优先考虑应用。
- CL指标亦可提供较高鉴别价值,但略低于LOC。
局限性
- 样本量较小(31个乳腺肿块),需更大样本验证。
- 研究基于既往相干波束合成数据,未在常规临床流程中前瞻性评估。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Traditional breast ultrasound imaging is a low-cost, real-time and portable method to assist with breast cancer screening and diagnosis, with particular benefits for patients with dense breast tissue. We previously demonstrated that incorporating coherence-based beamforming additionally improves the distinction of fluid-filled from solid breast masses, based on qualitative image interpretation by board-certified radiologists. However, variable sensitivity (range: 0.71-1.00 when detecting fluid-filled masses) was achieved by the individual radiologist readers. Therefore, we propose two objective coherence metrics, lag-one coherence (LOC) and coherence length (CL), to quantitatively determine the content of breast masses without requiring reader assessment. Data acquired from 31 breast masses were analyzed. Ideal separation (i.e., 1.00 sensitivity and specificity) was achieved between fluid-filled and solid breast masses based on the mean or median LOC value within each mass. When separated based on mean and median CL values, the sensitivity/specificity decreased to 1.00/0.95 and 0.92/0.89, respectively. The greatest sensitivity and specificity were achieved in dense, rather than non-dense, breast tissue. These results support the introduction of an objective, reader-independent method for automated diagnoses of cystic breast masses.
引用
- Wiacek, Oluyemi, Myers, et al. Coherence Metrics for Reader-Independent Differentiation of Cystic From Solid Breast Masses in Ultrasound Images. Ultrasound Med Biol. 2022. PMID: 36333154. doi:10.1016/j.ultrasmedbio.2022.08.018
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。