新型深度学习重建算法对超高分辨率CT图像质量的影响:临床与体模研究
结果显示New DLR整体图像质量评分显著更高(4.2±0.4 vs 3.3±0.4,p<0.0001),噪声功率谱√AUC更低(13.8 vs 14.2),重建时间中位数更短(5.0 vs 7.8 min)。结论认为新DLR在实用重建时间内提升了图像质量。
要点
- New DLR较Conv DLR整体图像质量评分更高,且所有New DLR图像评分在平均以上或优秀。
- New DLR的噪声功率谱√AUC更低,提示图像噪声降低。
- New DLR重建时间中位数5.0 min,短于Conv DLR的7.8 min。
核心结论
- 新型DLR算法在实用重建时间内改善了超高分辨率CT的图像质量。
- 新DLR使操作者可在提升图像质量与降低剂量之间进行选择。
研究设计与样本
- 图像分别采用New DLR(AiCE Body Sharp)和Conv DLR(AiCE Body)重建。
- 由两名放射科医师采用5分法(5=优秀,1=不可接受)评估整体图像质量。
- 计算噪声功率谱(NPS)及0.05–0.50 cycle/mm之间的√AUC NPS作为图像噪声指标。
- 采用Dunnett检验进行视觉评分统计分析,p<0.05为有统计学意义。
关键影像表现/征象
- New DLR整体图像质量评分为4.2±0.4,Conv DLR为3.3±0.4。
- 所有New DLR图像的整体图像质量评分均在平均以上或优秀。
- New DLR的√AUC NPS值为13.8,低于Conv DLR的14.2。
- New DLR与Conv DLR的重建时间中位数分别为5.0 min和7.8 min。
诊断标准与关键数值
- 整体图像质量采用5分法评估:5=优秀,1=不可接受。
- 图像噪声以0.05–0.50 cycle/mm之间的√AUC NPS表示,New DLR为13.8,Conv DLR为14.2。
临床应用与阅片要点
- 在超高分辨率CT增强检查中,可优先选用New DLR以同时获得更优图像质量和更短重建时间。
- New DLR的噪声降低特性可能有助于在维持图像质量的前提下探索降低辐射剂量的扫描方案。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To demonstrate the effect of an improved deep learning-based reconstruction (DLR) algorithm on Ultra-High-Resolution Computed Tomography (U-HRCT) scanners.
METHODS: Clinical and phantom studies were conducted. Thirty patients who underwent contrast-enhanced CT examination during the follow-up period were enrolled. Images were reconstructed using improved DLR [termed, New DLR, i.e., Advanced Intelligent Clear-IQ Engine (AiCE) Body Sharp] and conventional DLR (Conv DLR, AiCE Body) algorithms. Two radiologists assessed the overall image quality using a 5-point scale (5 = excellent; 1 = unacceptable). The noise power spectra (NPSs) were calculated to assess the frequency characteristics of the image noise, and the square root of area under the curve (√AUC NPS) between 0.05 and 0.50 cycle/mm was calculated as an indicator of the image noise. Dunnett's test was used for statistical analysis of the visual evaluation score, with statistical significance set at p < 0.05.
RESULTS: The overall image quality of New DLR was better than that of the Conv DLR (4.2 ± 0.4 and 3.3 ± 0.4, respectively; p < 0.0001). All New DLR images had an overall image quality score above the average or excellent. The √AUCNPS value of New DLR was lower than that of Conv DLR (13.8 and 14.2, respectively). The median values of reconstruction time required with New DLR and Conv DLR were 5.0 and 7.8 min, respectively.
CONCLUSIONS: The new DLR algorithm improved the image quality within a practical reconstruction time.
ADVANCES IN KNOWLEDGE: The new DLR enables us to choose whether to improve image quality or reduce the dose.
引用
- Sakai, Hida, Matsuura, et al. Impact of a new deep-learning-based reconstruction algorithm on image quality in ultra-high-resolution CT: clinical observational and phantom studies. Br J Radiol. 2022. PMID: 36318483. doi:10.1259/bjr.20220731
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。