连续剪切波弹性成像在病理髌腱中的可靠性
本研究纳入髌腱病患者、ACL重建术后患者及健康对照,评估连续剪切波弹性成像(cSWE)测量髌腱黏度和剪切模量的信度。结果显示隔日稳定性良好(黏度ICC=0.905,剪切模量ICC=0.805),但评分者间信度欠佳(ICC=0.591和0.532),建议由单一评分者进行测量。
要点
- cSWE测量髌腱黏度和剪切模量的隔日稳定性良好。
- 评分者间信度欠佳,临床测量应由同一评分者完成。
- cSWE可用于纵向量化髌腱黏弹性特征。
核心结论
- cSWE是量化髌腱黏弹性特征随时间变化的可靠工具。
- 因评分者间信度不理想,建议由单一评分者进行测量。
研究设计与样本
- 研究纳入髌腱病患者13例、骨-髌腱-骨自体肌腱ACL重建术后患者9例及无髌腱损伤史者13例。
- cSWE由多名评分者在2天内共进行4次测量,计算组内相关系数(ICC)和最小可检测变化(MDC95%)。
诊断标准与关键数值
- 隔日稳定性:黏度ICC=0.905,MDC95%=8.3 Pa·s;剪切模量ICC=0.805,MDC95%=27.4 kPa。
- 评分者间信度:黏度ICC=0.591,剪切模量ICC=0.532。
临床应用与阅片要点
- 在纵向随访中可采用cSWE量化髌腱黏弹性,但应由同一评分者完成测量以减少评分者间变异。
- 解释结果时可结合MDC95%判断变化是否超过测量误差(黏度8.3 Pa·s,剪切模量27.4 kPa)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Patellar tendon injuries occur via various mechanisms such as overuse, or due to surgical graft harvest for anterior cruciate ligament reconstruction (ACLR). Quantified patellar tendon stiffness after injury may help guide clinical care. Continuous shear wave elastography (cSWE) allows for the assessment of viscosity and shear modulus in tendons. The reliability of the measure, however, has not been established in the patellar tendon. The purpose of this study was to investigate the interrater reliability, intrarater reliability, and between-day stability of cSWE in both healthy and pathological patellar tendons.
METHODS: Participants with patellar tendinopathy (n = 13), history of ACLR using bone-patellar tendon-bone autograft (n = 9), and with no history of patellar tendon injury (n = 13) were recruited. cSWE was performed 4 times by multiple raters over 2 days. Intraclass correlations (ICC) and minimum detectable change (MDC95% ) were calculated.
RESULTS: Good to excellent between-day stability were found for viscosity (ICC = 0.905, MDC95% = 8.3 Pa seconds) and shear modulus (ICC = 0.805, MDC95% = 27.4 kPa). The interrater reliability measures, however, were not as reliable (ICC = 0.591 and 0.532).
CONCLUSIONS: cSWE is a reliable assessment tool for quantifying patellar tendon viscoelastic properties over time. It is recommended, however, that a single rater performs the measure as the interrater reliability was less than ideal.