肺结节半自动体积测量:标准剂量与胸片等效超低剂量CT的个体内比较
结果显示,标准剂量与超低剂量扫描间结节体积一致性高(ICC最高达0.999),但取决于重建核与迭代重建算法的选择;非完全相同重建的图像序列间比较需谨慎。
要点
- 超低剂量CT与标准剂量CT的结节体积一致性高,但需注意重建参数匹配。
- FBP Br40和ADMIRE-5 Bl64重建下标准与超低剂量间一致性最佳(ICC分别为0.995和0.994)。
- ADMIRE-3重建的扫描间一致性降低(ICC 0.788-0.882),建议对比时统一重建方案。
局限性
- 单中心研究
- 仅评估实性肺结节
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To assess the performance of semi-automated volumetry of solid pulmonary nodules on single-energy tin-filtered ultralow dose (ULD) chest CT scans at a radiation dose equivalent to chest X-ray relative to standard dose (SD) chest CT scans and assess the impact of kernel and iterative reconstruction selection.
METHODS: Ninety-four consecutive patients from a prospective single-center study were included and underwent clinically indicated SD chest CT (1.9 ± 0.8 mSv) and additional ULD chest CT (0.13 ± 0.01 mSv) in the same session. All scans were reconstructed with a soft tissue (Br40) and lung (Bl64) kernel as well as with Filtered Back Projection (FBP) and Iterative Reconstruction (ADMIRE-3 and ADMIRE-5). One hundred and forty-eight solid pulmonary nodules were identified and analysed by semi-automated volumetry on all reconstructions. Nodule volumes were compared amongst all reconstructions thereby focusing on the agreement between SD and ULD scans.
RESULTS: Nodule volumes ranged from 58.5 (28.8-126) mm3 for ADMIRE-5 Br40 ULD reconstructions to 72.5 (39-134) mm3 for FBP Bl64 SD reconstructions with significant differences between reconstructions (p < 0.001). Interscan agreement of volumes between two given reconstructions ranged from ICC = 0.605 to ICC = 0.999. Between SD and ULD scans, agreement of nodule volumes was highest for FBP Br40 (ICC = 0.995), FBP Bl64 (ICC = 0.939) and ADMIRE-5 Bl64 (ICC = 0.994) reconstructions. ADMIRE-3 reconstructions exhibited reduced interscan agreement of nodule volumes (ICCs from 0.788 - 0.882).
CONCLUSIONS: The interscan agreement of node volumes between SD and ULD is high depending on the choice of kernel and reconstruction algorithm. However, caution should be exercised when comparing two image series that were not identically reconstructed.