文献速递儿科影像腹部胸部MR

儿童体部成像的先进磁共振序列

State-of-the-art magnetic resonance imaging sequences for pediatric body imaging.
PubMed 文献 · Pediatr Radiol · 2022-10-18

该综述概述了用于儿童体部MRI的数据采样技术,包括并行成像、压缩感知、径向k空间采样、部分k空间采样和同时多层采集,并介绍了基于这些技术的新型T1、T2和DWI序列。文章还讨论了Dixon方法和减少金属伪影的新方法,旨在克服儿童体部MRI中检查时间长、需麻醉和不能屏气等限制。

要点

  • 径向k空间填充(如stack-of-stars和BLADE/PROPELLER)可减少运动伪影。
  • 部分k空间填充结合压缩感知等技术可实现自由呼吸、快速多期动态成像。
  • 同时多层采集可缩短DWI扫描时间并减少伪影。

关键影像表现/征象

  • 减少运动伪影的序列主要利用径向k空间填充,如T1加权成像的stack-of-stars方法和T2加权成像的PROPELLER/BLADE方法。
  • 动态成像中提高时间分辨率并在单次屏气内获得多期的序列主要使用某种形式的部分k空间填充方法。
  • 新序列采用压缩感知、黄金角径向k空间填充、并行成像和部分k空间填充等数据采样方法的可变组合,实现自由呼吸、快速序列。
  • 同时多层方法改进了扩散加权成像(DWI),减少了扫描时间和伪影。

临床应用与阅片要点

  • 在儿童体部MRI中,可优先考虑径向k空间填充序列以减少运动伪影。
  • 对于不能屏气的儿童,可采用结合压缩感知和部分k空间填充的自由呼吸快速序列。
  • DWI可选用同时多层采集以缩短扫描时间并减少伪影。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Longer examination time, need for anesthesia in smaller children and the inability of most children to hold their breath are major limitations of MRI in pediatric body imaging. Fortunately, with technical advances, many new and upcoming MRI sequences are overcoming these limitations. Advances in data acquisition and k-space sampling methods have enabled sequences with improved temporal and spatial resolution, and minimal artifacts. Sequences to minimize movement artifacts mainly utilize radial k-space filling, and examples include the stack-of-stars method for T1-weighted imaging and the periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER)/BLADE method for T2-weighted imaging. Similarly, the sequences with improved temporal resolution and the ability to obtain multiple phases in a single breath-hold in dynamic imaging mainly use some form of partial k-space filling method. New sequences use a variable combination of data sampling methods like compressed sensing, golden-angle radial k-space filling, parallel imaging and partial k-space filling to achieve free-breathing, faster sequences that could be useful for pediatric abdominal and thoracic imaging. Simultaneous multi-slice method has improved diffusion-weighted imaging (DWI) with reduction in scan time and artifacts. In this review, we provide an overview of data sampling methods like parallel imaging, compressed sensing, radial k-space sampling, partial k-space sampling and simultaneous multi-slice. This is followed by newer available and upcoming sequences for T1-, T2- and DWI based on these other advances. We also discuss the Dixon method and newer approaches to reducing metal artifacts.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. Kraus, Coblentz, Deshpande, et al. State-of-the-art magnetic resonance imaging sequences for pediatric body imaging. Pediatr Radiol. 2022. PMID: 36255456. doi:10.1007/s00247-022-05528-y

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。